WHO-prequalified IVDs: the CRP route into Malaysia.
Since 3 February 2026 MDA can register a WHO-prequalified in vitro diagnostic by relying on WHO’s own assessment. Here is who qualifies, the five steps in MDA’s flowchart, the 90-day review target, how changes are handled after registration, and the questions MDA has not yet answered.
What changed in 2026.
MDA became a participating national regulatory authority in the WHO Collaborative Registration Procedure (CRP) for WHO-prequalified in vitro diagnostics on 3 February 2026. With the manufacturer’s consent, WHO shares its prequalification dossier review, site inspection and performance evaluation reports with MDA, so MDA can verify the product is the same instead of assessing it again. MDA aims to finish its review within 90 working days of receiving everything from WHO. Participation is voluntary, Malaysian requirements and fees still apply, and MDA keeps the final decision.
WHO’s list of countries in the CRP for IVDs, dated 6 July 2026, has 43 entries. Malaysia, Thailand and the Philippines are the ASEAN members on it.
Who can use it
| MDA criterion | What it requires |
|---|---|
| Scope | Manufacturers or authorised representatives registering WHO-prequalified IVDs in Malaysia who choose to use the CRP |
| WHO prequalification status | The IVD is prequalified and listed under the WHO Prequalification Programme |
| Product sameness | Identical in all essential aspects to the WHO-prequalified version |
| Manufacturer consent | Written consent to WHO to share the relevant confidential assessment reports with MDA |
| Malaysian requirements | All applicable Malaysian requirements, submission procedures and fees |
WHO’s procedure (section 4.2) defines the same product as one with the same product name, regulatory version and product codes; the same manufacturing site and quality management system; the same quality, safety and performance data; the same design, with the same components from the same suppliers; and the same labelling, packaging, instructions for use and intended use. Labels and instructions may be translated if the content stays the same.
The five steps in MDA’s flowchart
- Confirm WHO prequalification. The IVD must be prequalified before the pathway starts. The technical information in the Malaysian dossier must be the same as that submitted to WHO.
- Give consent to WHO. The manufacturer sends WHO a signed consent form (Appendix 2 of the CRP procedure) authorising WHO to share its assessment reports with MDA for that product. It may go in with the expression of interest and the application; WHO and MDA share copies with each other.
- Submit an expression of interest (Appendix 3A) to MDA with the national registration application. If the applicant is not the manufacturer, as when a Malaysian authorised representative applies, WHO’s procedure also requires the manufacturer’s authorisation letter confirming the applicant acts for it.
- Submit the national registration application: the full dossier, in strict accordance with Malaysian requirements. For IVDs that means the CSDT format in MDA’s IVD guidance; see our CSDT page.
- MDA evaluates the submission and the declared differences. Any difference between the WHO dossier and the Malaysian one, such as local labelling or climatic stability data, must be declared.
Review time
| Source | Target | When the clock runs |
|---|---|---|
| MDA announcement (2026) | 90 working days | From a complete application and receipt of all relevant information from WHO; may vary with completeness, MDA queries and national administrative processes |
| WHO CRP procedure (Annex 4, 2021), section 4.6 | 90 calendar days of regulatory time | From a valid application or access to WHO’s information, whichever is later; excludes the applicant’s time to answer queries or fill gaps |
The two figures differ: 90 working days is roughly four months, 90 calendar days about three. We plan on MDA’s figure. Under WHO’s procedure, if an applicant fails to reply within a reasonable time (WHO gives 90 days as an example), MDA may end the CRP for that product and move it to normal registration.
After registration
- Changes go to both regulators. MDA requires post-registration changes to be submitted to WHO and MDA under section 6 of the CRP procedure. WHO asks applicants to submit changes that are subject to national rules within 30 calendar days of WHO accepting them. Translating labelling into the local language with no effect on quality, safety or performance is not reportable.
- Staying the same matters. If the Malaysian product stops matching the prequalified product, through a change made in one place and not the other, WHO removes it from the list of products registered under the procedure. The Malaysian change rules are on our change notification page.
- Withdrawals are shared. WHO tells MDA if a prequalified IVD is withdrawn, suspended or delisted, and MDA tells WHO if it suspends or cancels the Malaysian registration.
- Malaysian post-market duties still apply: problem reporting, recalls and FCAs, and re-registration every 5 years.
What MDA has not said
- Whether the CAB step is waived. Act 737 section 7 requires Class B, C and D devices to pass conformity assessment by a registered CAB. MDA’s announcement keeps all national requirements and does not mention CABs. Ask MDA before you skip it.
- Which MeDC@St option to use and where the expression of interest and consent form are uploaded. The announcement gives an enquiry email but no system guide.
- How many products have used it. MDA has published no figures.
CRP or verification?
| Route | Relies on | Published MDA time |
|---|---|---|
| WHO CRP (IVDs only) | WHO prequalification assessment, shared by WHO | 90 working days after WHO information received |
| Verification route | Approval by an EU notified body, US FDA, Health Canada, TGA, Japan, UK MHRA, Singapore HSA or Thai FDA, verified by a CAB | 30 working days after the CAB certificate and complete documents |
| Full conformity assessment | A CAB’s assessment of the full dossier | See our timeline page |
Many WHO-prequalified tests also hold an EU certificate. If it is under an annex MDA/GD/0070 accepts, verification has the shorter published MDA review. The CRP is the better fit when WHO prequalification is the product’s strongest approval. Malaysian IVD classes and dossier rules are on our IVD registration page.
Adakah Malaysia menyertai Prosedur Pendaftaran Kolaboratif (CRP) WHO?
Ya. Mulai 3 Februari 2026, MDA menyertai Prosedur Pendaftaran Kolaboratif WHO untuk peranti diagnostik in vitro (IVD) yang telah diprakelayakan oleh WHO. Pengilang perlu memberi kebenaran bertulis kepada WHO untuk berkongsi laporan penilaian dengan MDA, menghantar Pernyataan Minat (Appendix 3A) bersama permohonan pendaftaran, dan mengisytiharkan sebarang perbezaan seperti label tempatan. MDA menyasarkan semakan dalam 90 hari bekerja selepas menerima maklumat penuh daripada WHO. Keperluan dan fi nasional masih terpakai. WhatsApp 010-206 2070.
Frequently asked questions
Does Malaysia participate in the WHO Collaborative Registration Procedure?
Which products can use the CRP in Malaysia?
How long does MDA take under the CRP?
Is the CRP cheaper than normal MDA registration?
Does the CRP replace the conformity assessment body step?
What must be declared to MDA?
What happens when the product changes after registration?
Our IVD has WHO prequalification and a CE certificate. Which route should we use?
Registering a WHO-prequalified IVD?
Send us the WHO prequalification listing, the product codes and any CE, FDA or other approvals. We will compare the CRP with the verification route, act as your licensed Malaysian authorised representative, and prepare the dossier and declared differences.