IEC 60601 testing for MDA registration
If your device plugs in or runs on a battery, the CAB and MDA will look for electrical safety and EMC evidence. Here is what MDA asks for, which IEC 60601 standards apply, which editions are current, and where testing can be done in Malaysia.
Short answer: for active medical devices, MDA/GD/0070 asks for evidence of electrical safety, mechanical and environmental protection and electromagnetic compatibility, “e.g. summary and electrical safety test protocols, reports, certificates (IEC 60601-1 series)”, and for Class B, C and D the CSDT must include the electrical safety test data. The current general standard is IEC 60601-1 edition 3.2 (2005 + A1:2012 + A2:2020), and the EMC standard is IEC 60601-1-2:2014 + A1:2020, plus any particular standard for your device type. MDA publishes no list of recognised standards; where an older edition was used, it asks for a gap assessment.
What MDA asks for
| Where | What it says |
|---|---|
| MDA/GD/0070 (Second Edition, October 2025), Annex 3 (Class A submissions), active medical devices | Evidence supporting electrical safety, mechanical and environmental protection and EMC, with a summary of the non-clinical evidence: electrical safety test protocols, reports, certificates (IEC 60601-1 series) |
| MDA/GD/0070, Annex 4 (Class B, C and D) | Pre-clinical studies with the results of all verification and validation tests, including pre-clinical physical tests. Where outdated standards were applied, a gap assessment to show the state of the art |
| MDA/GD/0008 (CSDT guidance), 6.4.1 and 6.2 | Pre-clinical physical tests such as electrical safety tests, with complete data. Each standard identified by full title, number, date and issuing organisation; reports to give objectives, methodology, results, analysis and conclusions |
| Declaration of Conformity (MDA/GD/0070, section 10) | The vertical and horizontal standards applied must be stated |
The Essential Principles behind it
The Medical Device Regulations 2012, Third Schedule, Appendix 1:
- Paragraph 14, electrical risks: avoid, as far as possible, the risk of accidental electric shock during normal use and in single fault condition.
- Paragraph 13(6) and (7), EMC: reduce the risk of electromagnetic interference with other equipment, and give the device adequate intrinsic immunity to electromagnetic disturbance.
- Paragraph 13(1) and (4): programmable systems and software must be repeatable and reliable, and devices whose patient safety depends on external power need a power-failure alarm.
- Paragraphs 10(3) and 10(4): external influences such as magnetic fields, electrostatic discharge, temperature and humidity, and fire or explosion in normal use and single fault condition.
See our Essential Principles checklist guide for how the standards are cited against each principle.
Which standards: general, collateral and particular
IEC 60601-1 is the general standard. Applicable collateral standards (IEC 60601-1-x) apply together with it. Particular standards for a device type (IEC 60601-2-xx, IEC 80601-2-xx or ISO 80601-2-xx) can modify, replace or delete requirements, and a particular standard’s requirement takes priority (IEC 60601-1, clauses 1.3 and 1.4).
| Standard | Covers | Current edition (checked 11 November 2026) |
|---|---|---|
| IEC 60601-1 | General requirements for basic safety and essential performance | 3.2 (2005 + A1:2012 + A2:2020) |
| IEC 60601-1-2 | Electromagnetic disturbances (EMC) | 4.1 (2014 + A1:2020) |
| IEC 60601-1-6 | Usability, aligned with IEC 62366-1 | 3.2 (2010 + A1:2013 + A2:2020) |
| IEC 60601-1-8 | Alarm systems | 2.2 (2006 + A1:2012 + A2:2020) |
| IEC 60601-1-9 | Environmentally conscious design | 1.2 (2007 + A1:2013 + A2:2020) |
| IEC 60601-1-10 | Physiologic closed-loop controllers | 1.2 (2007 + A1:2013 + A2:2020); new edition in development |
| IEC 60601-1-11 | Home healthcare environment | 2.1 (2015 + A1:2020) |
| IEC 60601-1-12 | Emergency medical services environment | 1.1 (2014 + A1:2020) |
Particular standards include IEC 60601-2-24 (infusion pumps), IEC 80601-2-30 (automated blood pressure monitors) and IEC 80601-2-52:2026 (medical beds for adults, published May 2026). Software follows IEC 62304:2006 + A1:2015, and usability IEC 62366-1:2015 + A1:2020. Compliance with IEC 60601-1 also requires a risk management process to ISO 14971; see our ISO 14971 guide.
Terms you will see in a test report
- Basic safety: freedom from unacceptable risk directly caused by physical hazards in normal and single fault condition.
- Essential performance: the performance needed to keep the device free from unacceptable risk. The manufacturer defines it, and it drives what is tested.
- Applied parts, Type B, BF and CF: the parts that contact the patient, in rising order of protection against electric shock. CF is the strictest, for parts that may contact the heart.
- MOOP and MOPP: means of protection for operators and for patients. MOPP requirements are stricter.
- Class I and Class II: Class I relies on protective earthing as well as basic insulation; Class II uses double or reinforced insulation with no protective earth.
Which edition MDA expects
- No list, no cut-off date. The Regulations refer to standards “recognized by the Authority”, but MDA publishes no list, and its guidance names only the “IEC 60601-1 series”. MDA/GD/0070 asks for a gap assessment where outdated standards were used.
- Order of preference. Under MDA/GD/0031, CABs prefer Malaysian Standards and/or international (ISO, IEC, ITU) standards, then standards of the recognised countries (US, EU, Canada, Australia, Japan), then industry or company standards. A manufacturer that departs from this order must justify it.
- Malaysian adoptions lag. MS IEC 60601-1:2006 adopts the 2005 base edition without the 2012 and 2020 amendments, and MS IEC 60601-1-2:2007 adopts the 2001 EMC edition. Test to the current IEC editions.
- Fourth edition. IEC’s webstore forecasts IEC 60601-1 edition 4 for 2029. Until it is published, edition 3.2 is the state of the art.
Testing in Malaysia
| Laboratory | SAMM accreditation scope relevant here |
|---|---|
| SIRIM QAS International, Testing Services Department (SAMM 086) | Full type test to IEC 60601-1 editions 3.0, 3.1 and 3.2 and MS IEC 60601-1:2006, except clauses 9.4.2.4 and 15.3.5; EMC to IEC 60601-1-2:2014 + A1:2020, except clause 8.11; parts of IEC 60601-1-11 and IEC 80601-2-60 |
| SGS (Malaysia) Sdn Bhd, Senai (SAMM 382) | EMC to IEC 60601-1-2:2014 and its 2020 amendment |
From the schedules on Standards Malaysia’s SAMM directory, searched for “60601” on 11 November 2026. Overseas reports, including IECEE CB reports prepared for other markets, are commonly used; MDA has not published a rule on them either way. The draft Third Edition of MDA/GD/0070 says pre-clinical studies should be conducted in accredited laboratories; it is not final.
Energy Commission approval
Suruhanjaya Tenaga regulates listed electrical equipment under its Guidelines for the Approval of Electrical Equipment (2024 Edition). The list of 34 categories has no category for medical electrical equipment. It does include massagers, such as massage chairs and foot massagers, tested to MS IEC 60335. A wellness product that is not a medical device may therefore need ST approval as an electrical appliance rather than MDA registration; see our borderline products guide.
Type testing is not in-service testing
IEC 60601-1 tests the design. IEC 62353:2014 covers testing before equipment is put into service, during maintenance and after repair, and says it is not suitable for judging design compliance. In Malaysia, GDPMD paragraph 28 requires establishments supplying active devices to have installation, testing, commissioning and maintenance procedures in line with MS 2058, and section 43 of Act 737 makes the user ensure the device was properly installed, tested and commissioned. See testing and commissioning.
Our role
We review electrical safety and EMC reports, write gap assessments for older editions and put the evidence into the CSDT and Essential Principles checklist as part of our registration fee (Class A RM3,800, Class B RM6,800, Class C RM9,800, Class D RM12,000, with CAB and MDA fees). Testing itself is paid to the lab. See our MDA registration service.
Adakah MDA memerlukan ujian IEC 60601 untuk peranti perubatan aktif?
Ya, bagi peranti perubatan aktif. MDA/GD/0070 meminta bukti keselamatan elektrik, perlindungan mekanikal dan persekitaran serta keserasian elektromagnet (EMC), termasuk protokol, laporan dan sijil ujian (siri IEC 60601-1). Edisi semasa ialah IEC 60601-1 edisi 3.2 (2005 + A1:2012 + A2:2020) dan IEC 60601-1-2:2014 + A1:2020 bagi EMC. Jika edisi lama digunakan, MDA meminta penilaian jurang. Di Malaysia, SIRIM QAS International diakreditasi SAMM untuk ujian IEC 60601-1. WhatsApp 010-206 2070.
Frequently asked questions
Does MDA require IEC 60601 testing?
What is the current edition of IEC 60601-1?
Will MDA accept a test report to an older edition?
Does MDA accept CB test reports, or require an accredited lab?
Where can I get IEC 60601 testing done in Malaysia?
Do medical devices need Energy Commission (ST) approval?
Is IEC 62353 the same as IEC 60601?
Electrical safety evidence the CAB will accept
Tell us the device and what testing you have. We map the IEC 60601 standards that apply, check the editions and close the gaps before submission.