Sterilisation validation for MDA registration
Gloves, catheters, syringes, procedure packs and implants that leave the factory sterile need validation evidence in the MDA submission, whoever runs the steriliser. Here is what MDA asks for, how each method is validated, the current editions of the standards, and how MDA treats a change of method or steriliser.
Short answer: any device supplied sterile must have been sterilised by appropriate, validated methods (Medical Device Regulations 2012, Essential Principle 9(7)). For the MDA dossier, MDA/GD/0008 and MDA/GD/0070 ask for the initial sterilisation validation (bioburden, pyrogen and sterilant-residue testing, packaging validation), evidence of revalidation, the method, standards and sterility assurance level attained, a post-sterilisation functional test and, for ethylene oxide, residual data. MDA names no standard; the usual ones are ISO 11135 (EO), ISO 11137-1:2025 (radiation) and ISO 17665:2024 (steam). Class A sterile devices need the same evidence, even though they skip the CAB.
What MDA asks for on sterility
| Where | What it says |
|---|---|
| Medical Device Regulations 2012, Third Schedule, Appendix 1 (Essential Principles), paragraph 9(5)–(10) | A device labelled sterile must be sterilised by appropriate, validated methods and packed in a non-reusable pack so it stays sterile under the stated transport and storage conditions until the pack is damaged or opened; devices to be sterilised are made in controlled conditions; packaging and labels must tell sterile and non-sterile versions of the same product apart |
| MDA/GD/0008 (CSDT guidance), note to the design verification and validation section | For devices provided sterile: the initial sterilisation validation, including bioburden, pyrogen and sterilant-residue testing and packaging validation; evidence of ongoing revalidation of the packaging and sterilisation processes; method, SAL, standards, protocol and a summary of results; a post-sterilisation functional test; residual data for toxic sterilants such as EO |
| MDA/GD/0008, same note | If you rely on the validation of another device, justify why it applies and provide that device’s initial validation report. For a shelf life, data showing performance is kept until the expiry date |
| MDA/GD/0008, manufacturing information | A brief summary of the sterilisation method and processing |
| MDA/GD/0070 (Second Edition, October 2025), Annex 3 (Class A), Section 5 | “Supplied sterile”: sterilisation validation report and certificate, with the same contents as above |
| MDA/GD/0070, Annex 4 (Class B, C and D) | The CSDT’s pre-clinical evidence, and a label carrying the sterile or single-use symbol, expiry date, storage conditions and shelf-life information |
| MDA/GD/0031 (conformity assessment) | Laboratory tests such as sterility tests among the design verification evidence a CAB reviews; for Class A, which has no CAB audit, the manufacturer keeps a validation report as evidence of securing and maintaining sterile conditions |
Read from the Regulations, MDA/GD/0008, MDA/GD/0070, MDA/GD/0031 and MDA’s Class A submission requirements on 20 November 2026. None of them names a sterilisation standard; MDA/GD/0070 asks for a gap assessment where outdated standards were applied. Biocompatibility tests must be on the finished, sterilised device (MDA/GD/0008); see our ISO 10993 guide.
Methods and the standards that validate them
| Method | Standard | Typical use | Key validation elements | Residuals |
|---|---|---|---|---|
| Ethylene oxide (EO) | ISO 11135:2014 + Amd 1:2018 | Heat- and radiation-sensitive devices: plastics, catheters, tubing sets, procedure packs, devices with electronics | IQ/OQ/PQ; microbiological PQ (commonly an overkill half-cycle with biological indicators, or a bioburden-based approach); physical PQ of temperature, humidity and gas concentration; load configuration | EO and ethylene chlorohydrin, to ISO 10993-7:2026; aeration validated |
| Gamma (cobalt-60) | ISO 11137-1:2025, dose setting ISO 11137-2, dosimetry ISO 11137-3:2017 | Single-use plastics, gloves, syringes, sutures, many implants, where the materials tolerate the dose | Maximum acceptable dose for the materials; sterilisation dose set or a 25 or 15 kGy dose substantiated; dose mapping; routine dosimetry; periodic dose audits | No chemical residues; check material and packaging degradation |
| Electron beam | ISO 11137 series | Low-density, uniformly packed products; short exposure times | As for gamma, with dose mapping for the lower penetration of electrons | None; check material effects |
| X-ray | ISO 11137 series | Dense or mixed loads; an alternative to cobalt-60 capacity | As for gamma: dose setting, dose mapping, routine dosimetry, dose audits | None; check material effects |
| Moist heat (steam) | ISO 17665:2024 | Heat- and moisture-tolerant devices: metal instruments, some liquids, reusable devices | IQ/OQ/PQ; temperature and pressure mapping of the load; biological indicators where relevant | None; check water and heat effects |
| Vaporised hydrogen peroxide (VH2O2) | ISO 22441:2022 | Low-temperature processing of heat-sensitive devices, in hospitals and increasingly in industry | Development, validation and routine monitoring and control as ISO 22441 requires | Assess any remaining hydrogen peroxide in the biological evaluation |
Other sterilising agents are validated under ISO 14937:2009, the general standard. ISO 11137-1 covers radiation from cobalt-60 or caesium-137, electron beams and X-rays. Contract capacity exists in Malaysia: for example, STERIS AST lists EO, gamma and X-ray processing with laboratory testing at Kuala Ketil, Kedah, and gamma at Rawang, Selangor.
How a sterilisation process is validated
- Product and process definition. Product families and the worst-case product, the materials’ tolerance to the process (for radiation, the maximum acceptable dose), and the sterile barrier system.
- Bioburden. Count the micro-organisms on product before sterilisation, under ISO 11737-1 (2018 + Amendment 1:2021). Bioburden drives radiation dose setting and EO bioburden-based approaches, and MDA asks for it explicitly.
- Installation and operational qualification (IQ/OQ) of the steriliser or irradiator, usually done by the contract steriliser for its own equipment.
- Performance qualification (PQ). Physical PQ (temperature, humidity and gas for EO; dose mapping for radiation; temperature mapping for steam) and microbiological PQ, including tests of sterility under ISO 11737-2:2019 on product exposed to a reduced treatment.
- Release and routine control. Each load released on biological indicators, dosimetry or parametric release, as validated.
- Maintaining the validation. Requalification at defined intervals and after changes; for radiation, sterilisation dose audits under ISO 11137-2 to show the dose is still effective. MDA asks for this as “evidence of the on-going revalidation”.
The target is a sterility assurance level of 10⁻⁶ or better for a terminally sterilised device labelled STERILE: EN 556-1 sets that probability of a viable micro-organism, and ISO 11137-2 gives the methods that substantiate 25 kGy or 15 kGy as achieving it. Sterilisation validation sits inside your risk management file; see ISO 14971.
Packaging and shelf life
- Sterile barrier system. ISO 11607-1 (materials, sterile barrier and packaging systems) and ISO 11607-2 (forming, sealing and assembly processes), both 2019 with Amendment 1:2023 on risk management, both reconfirmed by ISO in 2024. MDA lists packaging validation as part of the initial sterilisation validation.
- Shelf life. MDA/GD/0008 accepts accelerated ageing validated with real-time data, real-time studies, or both. Accelerated data can support the first claim if the parameters are justified, but must be backed by real-time data. ASTM F1980-21 is the usual guide for accelerated ageing.
- Label and instructions. The expiry date as at least year and month for sterile single-use devices, an indication that the device is sterile, information on the sterilisation method, and what to do if the sterile pack is damaged (Sixth Schedule of the Regulations; MDA/GD/0026, 7th Edition). See labelling.
Current editions of the standards you will cite
| Standard | Covers | Status on 20 November 2026 |
|---|---|---|
| ISO 11135 | Ethylene oxide | 2014 + Amd 1:2018 current; Edition 3 closed its final (FDIS) vote on 12 November 2026 |
| ISO 11137-1 | Radiation: requirements | 2025 (Edition 2, April 2025), replacing 2006 and its 2013 and 2018 amendments |
| ISO 11137-2 | Radiation: establishing the dose, dose audits | 2013 + Amd 1:2022 current; Edition 4 draft vote closed 7 November 2026 |
| ISO 17665 | Moist heat | 2024 (Edition 1, March 2024), replacing ISO 17665-1:2006 and the -2 and -3 technical specifications |
| ISO 22441 | Low-temperature vaporised hydrogen peroxide | 2022 (Edition 1) |
| ISO 14937 | Any other sterilising agent | 2009, reconfirmed 2026 |
| ISO 11737-1 / -2 | Bioburden; tests of sterility in validation | 2018 + Amd 1:2021 (revision decided); 2019, reconfirmed 2025 |
| ISO 11607-1 / -2 | Packaging for terminally sterilised devices | 2019 + Amd 1:2023, reconfirmed 2024 |
| ISO 10993-7 | EO residuals | 2026 (Edition 3, April 2026) |
From ISO’s committee pages, 20 November 2026; ISO 10993-7 as checked for our biocompatibility page on 11 November 2026. If your validation cites a withdrawn edition, such as ISO 11137-1:2006 or ISO 17665-1:2006, MDA/GD/0070 asks for a gap assessment rather than automatic revalidation.
After registration: changing the method, the steriliser or the pack
- Now (MDA/GD/0020, 4th Edition): any change to a sterilisation process, including an outsourced one, is Category 2 and needs MDA endorsement before you implement it. MDA’s examples: EO to gamma, new moist heat parameters, batch release to parametric release. A change of sterilisation facility with no process change is also listed in Category 2. See change notification.
- Next (MDA/GD/0072, May 2026, not yet in force): its Flowchart C treats changes that raise bioburden, make the device harder to sterilise, change the method or process, or change sterile primary packaging as significant. Renaming a sterilisation site, or adding one that runs an exact duplicate of the validated process, becomes a notification. It also lists suppliers of sterilisation services as critical suppliers.
How we help
- Gap assessment of your sterilisation, packaging and shelf-life evidence against MDA/GD/0008, MDA/GD/0070 and the current standards
- Protocol and report review for EO, radiation or steam validations before they run, so the data answers what MDA will ask
- Contract steriliser oversight: supplier controls, quality agreements and requalification records under your ISO 13485 system
- The sterility section of the CSDT, tied to the risk file and the biological evaluation, then the MDA submission
- Change notifications when you move steriliser, method or packaging
Writing the sterility section is part of our registration fee when we register the device: Class A RM3,800, Class B RM6,800, Class C RM9,800, Class D RM12,000, all-in with CAB and MDA fees. Validation runs and laboratory tests are paid to the steriliser and the lab. A standalone gap assessment depends on how many products and processes you have, so we quote it after seeing your reports. Malaysian factories setting up sterile production can start with our manufacturing consultant page (ISO 13485 and the MDA manufacturer licence, RM7,500 our fee). MDA Academy ran its own one-day course on sterilisation methods, validation and regulatory compliance on 6 August 2026; see our MDA training page for its calendar. Full dossier structure: CSDT technical file; the registration itself: MDA registration.
Adakah MDA memerlukan pengesahan pensterilan untuk peranti steril?
Ya. Peranti yang dibekalkan steril mesti disteril dengan kaedah yang sesuai dan disahkan (Prinsip Penting 9(7), Peraturan-Peraturan Peranti Perubatan 2012). MDA/GD/0008 dan MDA/GD/0070 meminta laporan pengesahan pensterilan awal (bioburden, pirogen, sisa bahan pensteril dan pengesahan pembungkusan), bukti pengesahan semula, kaedah, piawaian dan tahap jaminan kesterilan (SAL), ujian fungsi selepas pensterilan dan data sisa etilena oksida. Peranti Kelas A yang steril juga perlu bukti ini. Piawaian biasa: ISO 11135 (EO), ISO 11137-1:2025 (sinaran) dan ISO 17665:2024 (stim). WhatsApp 010-206 2070.
Frequently asked questions
Does MDA require sterilisation validation?
Which standard applies: ISO 11135, ISO 11137 or ISO 17665?
What sterility assurance level does MDA expect?
Can we rely on a validation done for another device or product family?
Do sterile Class A devices need the same evidence?
We are moving from EO to gamma, or to a new contract steriliser. What do we tell MDA?
Does sterilisation affect the biocompatibility evidence?
How do we support the shelf life on the label?
A sterility section MDA will accept
Tell us the device, the method and who sterilises it. We check the validation, packaging and shelf-life evidence and write the sterility section of the CSDT for MDA.