ISO 14971 risk management for MDA registration

Every Class B, C and D application to MDA carries a risk management report, and MDA’s submission guide names ISO 14971:2019 as the standard. Here is what MDA asks for, where it comes from in the law, and how the standard’s process maps onto it.

Short answer: for Class B, C and D, MDA/GD/0070 (Second Edition, October 2025) asks for “results of the risk analysis process conducted in accordance with ISO 14971:2019”, including the latest risk management report. The legal basis is the Medical Device Regulations 2012: the Essential Principles require risks to be controlled until acceptable against the benefits, and the CSDT must contain a risk management report. ISO 14971:2019 is still the current edition, confirmed by ISO in 2025. Class A does not upload the report, but must have it on file.

What MDA asks for

MDA/GD/0070 Annex 4, for Class B, C and D on the verification route:

MDA expectsWhere it comes from in ISO 14971:2019
The latest risk management reportClause 9, risk management review
A list of possible hazards of the deviceClause 5.4, hazards and hazardous situations
The technique used to analyse riskClause 4.4 (plan) and 5 (analysis): FMEA, FTA, PHA and similar
The evaluation of the risks against the claimed benefitsClauses 6, 7.4 and 8
The methods used to control or reduce risk to acceptable levelsClause 7, risk control
Who carried out the analysisClause 4.3, competence of personnel
For re-registration: an updated risk analysisClause 10, production and post-production

The same list appears in MDA’s MeDC@St requirements for Class B, C and D new and re-registration applications. On the full conformity assessment route, the CAB checklist in MDA/GD/0031 (Annex A) checks that the “risk management report demonstrated conformance with ISO 14971”.

Where it comes from in Malaysian law

  • Essential Principles (Medical Device Regulations 2012, Third Schedule, Appendix 1). Paragraph 2 accepts risks only when “acceptable risks when weighed against the benefits to the patient”. Paragraph 3(2) sets the order of control: identify hazards and estimate the risks from intended use and foreseeable misuse, eliminate risks as far as reasonably practicable by safe design, reduce what remains with protective measures, including alarms, and inform users of residual risks. Paragraph 7 requires the benefits to outweigh undesirable side effects.
  • The CSDT (Third Schedule, Appendix 2, paragraph 10). The CSDT “shall contain documentation on risk analysis”, “in the form of a risk management report”, based on “standards recognized by the Authority” and appropriate to the complexity and risk class of the device.
  • MDA/GD/0008, the CSDT guidance (First Edition, March 2014): “It is recommended that the risk management activities be conducted according to ISO 14971”, with the risk management plan and the results of risk assessment and risk control provided.

The Regulations define a recognised standard as one MDA deems to give a presumption of conformity with specific Essential Principles, but MDA publishes no list of recognised standards. The submission guides are where MDA names editions. See our Essential Principles checklist guide for how the standard is cited in the checklist.

The ISO 14971:2019 process, clause by clause

ClauseTitleWhat it requires
4General requirementsA documented risk management process, management responsibilities and policy for risk acceptability, competent people, a risk management plan for the device and a risk management file.
5Risk analysisIntended use and reasonably foreseeable misuse, characteristics related to safety, the hazards and hazardous situations, and an estimate of each risk.
6Risk evaluationEach risk compared with the acceptability criteria in the plan.
7Risk controlControl options, implementation and verification, residual risk evaluation, benefit–risk analysis where a risk is not acceptable, new risks created by the controls, and completeness.
8Overall residual riskThe combined residual risk judged against the benefits of the intended use.
9Risk management reviewA review before commercial release that the plan was carried out; the result is documented as the risk management report.
10Production and post-productionCollecting and reviewing information from production, complaints, incidents and the state of the art, and acting on it.

Clause titles from ISO’s Online Browsing Platform; the descriptions are ours. The standard requires objective criteria for risk acceptability but does not set acceptable risk levels; the manufacturer does that in its policy and plan.

Key terms

  • Harm: injury or damage to the health of people, or damage to property or the environment.
  • Hazard: a potential source of harm.
  • Hazardous situation: a circumstance in which people, property or the environment are exposed to one or more hazards.
  • Risk: the combination of the probability of occurrence of harm and the severity of that harm. The Malaysian Regulations use the same definition.
  • Residual risk: the risk remaining after risk control measures have been implemented.
  • Benefit: a positive impact or desirable outcome of using the device on an individual’s health, or on patient management or public health.

What changed from the 2007 edition

ISO 14971:2019 (third edition) replaced ISO 14971:2007. From its foreword:

  • Definitions of benefit, reasonably foreseeable misuse and state of the art were added.
  • Benefit–risk analysis terminology was aligned with the regulations.
  • Risks from data and systems security are explicitly in scope, alongside biocompatibility, electricity, moving parts, radiation and usability.
  • The results of the review are documented as the risk management report.
  • Production and post-production requirements were clarified and restructured.
  • Several informative annexes moved to the guidance document ISO/TR 24971:2020.

If your file was written to the 2007 edition, MDA/GD/0070 asks for a gap assessment where outdated standards were applied, to show the state of the art. The Malaysian Standard MS ISO 14971:2009 adopts the 2007 edition and has not been updated, so do not rely on it for a new file.

Guidance and versions you will meet

DocumentStatus (checked 11 November 2026)
ISO 14971:2019Current. Reviewed and confirmed by ISO in 2025; no amendment
EN ISO 14971:2019 + A11:2021Harmonised under the EU MDR from 17 May 2022. A11 adds the Annex Z tables only
ISO/TR 24971:2020Guidance on applying ISO 14971. Marked for revision in July 2026, to be replaced by ISO/TS 24971-1
MS ISO 14971:2009Malaysian adoption of ISO 14971:2007, confirmed 2015. No adoption of 2019

Where risk management shows up after approval

  • Post-market surveillance. Complaints, incidents and recalls feed clause 10. See post-market surveillance and mandatory problem reporting.
  • Changes. MDA’s change management guidance, MDA/GD/0072 (First Edition, May 2026, not yet in force), lists a risk management report among the evidence for significant changes such as a change of material in a body-contacting device. See change notification.
  • Re-registration every five years needs an updated risk analysis. See registration renewal.
  • Biocompatibility is evaluated inside the risk process too: ISO 10993-1 is written as a risk management exercise under ISO 14971. See our ISO 10993 guide.

MDA itself ran a course on ISO 14971:2019 risk management for medical devices at its Cyberjaya office on 13 August 2026, which tells you how much weight its reviewers put on it.

Our role

We prepare or review the risk management file as part of the CSDT, which is included in our registration fee (Class B RM6,800, Class C RM9,800, Class D RM12,000, with CAB and MDA fees). For a standalone review of an existing file, send it on WhatsApp for a quote. See our MDA registration service.

Jawapan ringkas · Bahasa Malaysia

Adakah MDA memerlukan ISO 14971 untuk pendaftaran peranti perubatan?

Ya, bagi Kelas B, C dan D. MDA/GD/0070 (Edisi Kedua, Oktober 2025) meminta keputusan analisis risiko yang dijalankan mengikut ISO 14971:2019, termasuk laporan pengurusan risiko terkini, senarai bahaya, teknik analisis, penilaian risiko berbanding manfaat, kaedah kawalan risiko dan pihak yang menjalankan analisis. Peraturan-Peraturan Peranti Perubatan 2012 mewajibkan laporan pengurusan risiko dalam CSDT. Kelas A tidak perlu memuat naik laporan ini, tetapi ia mesti disimpan dalam dokumentasi teknikal. WhatsApp 010-206 2070.

FAQ

Frequently asked questions

Does MDA require ISO 14971?
For Class B, C and D on the verification route, yes in practice. MDA/GD/0070 (Second Edition, October 2025) asks for the results of the risk analysis process conducted in accordance with ISO 14971:2019, including the latest risk management report. The Medical Device Regulations 2012 require a risk management report in the CSDT, based on standards recognised by the Authority, and MDA’s CSDT guidance MDA/GD/0008 recommends ISO 14971.
Which edition of ISO 14971 does MDA accept?
MDA’s registration submission guides name ISO 14971:2019, and the draft Third Edition of MDA/GD/0070 published for comment in October 2026 keeps it. ISO 14971:2019 is the current edition: ISO reviewed and confirmed it in 2025, and there is no amendment. If a file still follows the 2007 edition, MDA/GD/0070 asks for a gap assessment to show the state of the art.
Is EN ISO 14971:2019 + A11:2021 the same as ISO 14971:2019?
Yes for the requirements. ISO/TC 210 states that amendment A11 is specific to the EU: it adds Annexes Z mapping the standard to the EU regulations and does not modify the normative part of ISO 14971. A risk management file built to EN ISO 14971:2019 + A11:2021 for CE marking meets the ISO 14971:2019 requirement MDA names.
Do Class A devices need a risk management report?
It is not uploaded. MDA’s Class A submission requirements list neither a CSDT nor a risk management report. But the manufacturer must still keep a summary of technical documentation in CSDT format, available to MDA on request, and the Declaration of Conformity states that the device complies with all the Essential Principles, which include risk management. So the file must exist.
What must the risk management report contain for MDA?
MDA/GD/0070 Annex 4 lists: the latest risk management report, a list of possible hazards, the technique used to analyse risk, the evaluation of the risks against the claimed benefits of the device, the methods used to control or reduce risk to acceptable levels, and the individual or organisation that carried out the analysis. For re-registration, an updated risk analysis is required.
Is there a Malaysian Standard for ISO 14971?
Yes, but it is old. The Department of Standards Malaysia catalogue lists MS ISO 14971:2009 (confirmed 2015), an identical adoption of ISO 14971:2007. There is no MS adoption of the 2019 edition. MDA’s submission guides name ISO 14971:2019, so work to that.
Does ISO 14971 require ISO 13485?
No. ISO 14971 states that it does not require the manufacturer to have a quality management system, although risk management can be part of one. ISO 13485 in turn requires risk-based processes and product risk management, and MDA requires a quality management system for registration, so in practice the two run together.

Get the risk file right the first time

The risk management report is read by the CAB and by MDA, and gaps in it come back as queries. We build or fix it against ISO 14971:2019 and MDA’s Annex 4 list.