UDI in Malaysia: where MDA stands

Unique Device Identification is mandatory in the US and EU and being phased in across Asia. In Malaysia it is still voluntary, but MDA has a proposed timeline and a draft rule that would put UDIs into every distributor’s records.

Short answer: UDI is not mandatory in Malaysia today. MDA’s labelling guidance mentions it only as an example of e-labelling. MDA’s proposed timeline makes it voluntary for all classes in 2026, mandatory for Class D in 2027 and other classes in 2028, with full compliance by 2030, “once decided and assuming MeDC@St 3.0+ is completed”. The nearest concrete step is a draft distribution records guidance (comments close 18 November 2026) that would make establishments record the UDI where a device has one.

Malaysia’s position, document by document

DocumentWhat it says about UDIBinding?
Act 737, Medical Device Regulations 2012, Duties and Obligations Regulations 2019No UDI provision. Labels and records rely on batch, lot or serial numbersLaw
MDA/GD/0026 Requirements of Labelling, Seventh Edition (May 2026)UDI listed as an example of electronic labelling (“serves as a unique identifier for traceability”), next to barcodes and RFID. The label must carry a batch, lot or serial number for tracing and recallGuidance; UDI optional
MDA pilot survey announcement (July 2024)Pilot survey of Class D establishments, then Class B, in 2024, to decide the timeline and the updates to MeDC@St 3.0+. Published the proposed timeline belowNo
MDA Chief Executive, GS1 Malaysia Summit (September 2025)Repeated the timeline; said MDA announced UDI implementation to industry representatives on 8 July 2025; issuing agencies to be those recognised by MDANo
Draft Distribution Records, Second Edition (comment period 3–18 November 2026)Defines UDI (UDI-DI plus UDI-PI) and would require it in distribution records “where applicable”Draft

MDA’s proposed timeline

YearPhase
2024Pilot survey (done: Class D, then Class B establishments)
2026Transition: voluntary implementation for all classes
2027Mandatory phase 1: Class D
2028Mandatory phase 2: all classes other than D
2030Full UDI compliance in Malaysia

MDA labels this “Proposed UDI Timeline”, applying “once decided and assuming MeDC@St 3.0+ is completed”. MDA’s MeDC@St 3.0 plan, updated June 2024, puts the implementation phase in 2026 and all modules ready in 2027. No gazetted rule, circular or final guidance sets these dates yet, and MDA has not published the pilot survey results. We will update this page when it does.

The draft distribution records rule

Every establishment must keep distribution records under section 37 of Act 737. MDA’s draft Second Edition of the guidance (replacing MDA/GD/0012, First Edition, June 2020, which does not mention UDI) would require:

  • the device name, brand name, UDI and quantity in each record;
  • for devices without a UDI, the model or version and the batch or lot number, plus the MDA registration number and its validity period;
  • for implantable devices, the healthcare facility and department, patient details, implant date and explant details, so the device can be traced to the patient.

It is a draft and the final text may change. If you hold an establishment licence, your GDPMD records procedure is where this lands. See our GDPMD page and importing guide.

What a UDI is

The definitions MDA’s draft uses come from the IMDRF UDI guidance:

  • UDI: numeric or alphanumeric characters created through a globally accepted device identification and coding standard, made of a UDI-DI and a UDI-PI. “Unique” does not imply serialisation.
  • UDI-DI (device identifier): specific to a device model and the key to its record in a UDI database. Examples are a GS1 GTIN, an HIBC-LIC code or an ISBT 128-PPIC.
  • UDI-PI (production identifier): identifies the production unit: serial number, lot or batch number, software version, and manufacturing or expiry date.
  • UDI carrier: how the UDI is put on the label, by automatic identification and data capture (AIDC, such as a barcode or RFID) and, where applicable, a human-readable interpretation (HRI).

A GS1 UDI uses application identifiers: (01) the GTIN as the device identifier, (17) the expiry date, (10) the lot or batch and (21) the serial number. MDA’s own recall listings already quote UDIs in this form, for example (01)00889024379718(17)301014(10)85390029.

UDI in other markets

MarketStatus
United States (FDA)In force. Label UDI dates ran from Class III in 2014 to Class I in 2018, with enforcement for most Class I devices from September 2022. The GUDID database holds the device identifier only
European Union (MDR)UDI carriers on labels from 26 May 2021 (implantable and Class III), 26 May 2023 (Class IIa and IIb) and 26 May 2025 (Class I). The Basic UDI-DI appears on the EU declaration of conformity. EUDAMED’s UDI and device registration module became mandatory on 28 May 2026
Singapore (HSA)Phased: high-risk implants from 1 November 2022, Class D from 1 November 2024, Class C from 1 November 2026, Class B from 1 November 2028; Class A voluntary. HSA accepts UDIs applied for the EU or US as they are
China (NMPA)Already required for Class III and many Class II devices. A March 2026 announcement extends it to all Class II devices and Class I IVD reagents produced from 1 June 2027, and all Class I devices from 1 June 2029

From each regulator’s published guidance, checked 11 November 2026. HSA describes its phase dates as tentative. The China dates come from a provincial copy of the national announcement.

How to get ready

  1. If you already sell in the EU, US or Singapore, you have UDIs. Keep using them; a UDI is assigned per device, not per country.
  2. If you have none, get device identifiers from an issuing agency. GS1 Malaysia is the sole distributor of GS1 numbers here; its standard licence covers up to 1,000 GTIN-13 numbers, with a RM1,050 one-off registration fee and an annual fee from RM816, by issued capital.
  3. Keep the label compliant now. MDA/GD/0026 still requires a batch, lot or serial number for tracing and recall, and the MDA registration number. See labelling requirements.
  4. Match UDIs to your MDA registrations. Expect MeDC@St to ask for UDI data per registered device, and register families and groups consistently. See grouping.
  5. Update your distribution records procedure once MDA finalises the draft, especially for implantable devices.
Jawapan ringkas · Bahasa Malaysia

Adakah UDI wajib di Malaysia?

Belum. Tiada akta, peraturan atau garis panduan muktamad MDA yang mewajibkan UDI pada label. MDA/GD/0026 (Edisi Ketujuh, Mei 2026) hanya menyebut UDI sebagai contoh pelabelan elektronik. Garis masa cadangan MDA: sukarela bagi semua kelas pada 2026, wajib bagi Kelas D pada 2027 dan kelas lain pada 2028, pematuhan penuh pada 2030, tertakluk kepada keputusan MDA dan siapnya MeDC@St 3.0+. Draf Edisi Kedua garis panduan rekod pengedaran (komen 3–18 November 2026) akan mewajibkan rekod UDI jika ada. WhatsApp 010-206 2070.

FAQ

Frequently asked questions

Is UDI mandatory in Malaysia?
Not yet. No Act, regulation, circular or final MDA guidance requires a UDI on the label or a UDI submission. MDA’s labelling guidance, MDA/GD/0026 (Seventh Edition, May 2026), mentions UDI only as an example of electronic labelling. MDA has published a proposed timeline with mandatory phases from 2027, but it is labelled as proposed and depends on MDA’s new MeDC@St 3.0+ system being completed.
What is MDA’s UDI timeline?
MDA’s proposed timeline, first published with its July 2024 pilot survey and repeated by MDA’s Chief Executive in September 2025: pilot survey in 2024, a voluntary transition phase for all classes in 2026, Class D mandatory in 2027, other classes mandatory in 2028, and full UDI compliance by 2030. MDA marks it as applying “once decided and assuming MeDC@St 3.0+ is completed”.
What would the draft distribution records guidance change?
MDA published a draft Second Edition of its distribution records guidance (to replace MDA/GD/0012) for comment from 3 to 18 November 2026. It would require establishments to record the UDI, where applicable, with each device’s name, brand and quantity. For devices without a UDI, records would hold the model or version and the batch or lot number, plus the MDA registration number and its validity. Implantable devices would have to be traceable to the patient. It is a draft and may change.
What is the difference between UDI-DI and UDI-PI?
The UDI-DI identifies the device model and is the key to the device’s record in a UDI database; with GS1 it is a GTIN. The UDI-PI identifies the production unit: lot or batch number, serial number, software version, and manufacturing or expiry date. Together they form the UDI, which goes on the label as a barcode or other machine-readable carrier (AIDC) with a human-readable version (HRI).
Where do I get a UDI in Malaysia?
From an issuing agency. Internationally the main ones are GS1, HIBCC and ICCBBA (accredited by the US FDA) and, in the EU, also IFA. GS1 Malaysia is the sole distributor of GS1 numbers in Malaysia; its standard licence covers up to 1,000 GTIN-13 numbers, with a one-off registration fee of RM1,050 and an annual fee from RM816, depending on the company’s issued capital. MDA has said issuing agencies will be those it recognises but has not published a list.
Will a UDI from the EU or US work in Malaysia?
Probably, but MDA has not said. A UDI is assigned by the manufacturer once per device and is meant to be global; Singapore’s HSA, for example, accepts UDIs applied for the EU or US markets as they are. Until MDA publishes its requirements, keep the UDI you already use and make sure your Malaysian labelling still carries the lot or serial number MDA/GD/0026 requires.

Labels and records that will not need redoing