Skip to main content
Act 737 · MeDC@St 2.0+ · Electronic certificates · MMDR public search

MDA certificates: what each one proves, how to get it and how to check it.

“MDA certificate” can mean six different documents. Buyers, hospitals, tender boards, online platforms and foreign regulators each ask for a different one. Here is every certificate the Medical Device Authority issues, how long it lasts, where to download it and how to check one you have been given.

01 MDA Registration02 GDPMD + Licensing03 ISO 13485 QMS04 MDSAP05 CE / FDA Export

Short answer: an MDA certificate is usually the medical device registration certificate, issued for each device under section 7(1) of Act 737 with its registration number. It is valid for 5 years and has been electronic only since 1 November 2023, downloaded from MeDC@St 2.0+. The other one people mean is the company’s establishment licence (one per role, valid 3 years). Check either in MDA’s public MMDR search. A GDPMD or ISO 13485 certificate is not issued by MDA and does not let you sell a device.

Checked against MDA on 15 November 2026

Every certificate MDA issues

Fees for each licence and approval are in our medical device licensing table. This table is about the documents themselves.

CertificateWhat it provesWho holds itValid forFormatHow to check
Medical device registration certificateThe device is registered under Act 737 and has a registration number (s.7(1))The registration holder: the local manufacturer, or the authorised representative of a foreign manufacturer5 years from registration (reg. 6(3))Electronic only since 1 Nov 2023; download from MeDC@St 2.0+MMDR: Registered Medical Device
Establishment licenceThe company may act in one role: manufacturer, authorised representative, importer or distributor (s.15); one licence per role since 1 Jul 2024The company3 years from issue and from each renewal (regs. 12(3), 13(5))Electronic from 5 Jun 2024; electronic only for approvals after 15 Apr 2026MMDR: Licensed Establishment
Certificate of Free Sale / Manufacturing CertificateA registered device is lawfully sold in, or manufactured in, Malaysia (for importing regulators abroad)The exporter or manufacturer2 or 5 years, chosen at paymentElectronic by email since 1 Mar 2026Ask MDA’s International and Industry Affairs Division
Export CertificateA device made or imported only for export is exempt under the Exemption Order 2024The exporter2 or 5 years, chosen at paymentElectronic by email since 1 Mar 2026Ask MDA’s International and Industry Affairs Division
Product classification letterWhether a product is a medical device at all (it does not assign a risk class)The applicant5 years for letters issued from 1 Oct 2026By email from MDA’s online classification system since 1 Aug 2026The letter itself
Advertisement approvalAn advert for a registered device was approved; the ad carries the approval numberThe advertiserUntil the device registration expiresApproval with an approval numberMDA’s published list of advertisement approvals
Exemption letters (special access, export only, demonstration, custom-made)A named device or use is exempt under the Medical Device (Exemption) Order 2024The applicant named for each routeBy route: special access 6 months, export-only 5 yearsLetter from MDAMMDR: Special Access, Custom Made, Demonstration tabs

Not MDA certificates: the GDPMD certificate (issued by a conformity assessment body after a distribution audit), the ISO 13485 certificate (issued by a certification body) and a CAB’s conformity assessment report. They are inputs to an MDA licence or registration, not substitutes for one.

The medical device registration certificate

Section 7(1) of the Medical Device Act 2012 (Act 737) says that once a device has been through conformity assessment by a CAB and the fee is paid, MDA may register the device for a prescribed period, assign it a registration number and issue the applicant a certificate of registration. Regulation 6(3) of the Medical Device Regulations 2012 sets the period: the device stays on the Register for five years from the date of registration unless MDA cancels it.

Electronic since 2023

  • 2 May 2023: MDA began issuing electronic registration certificates, downloadable from MeDC@St 2.0+ once the application is complete and the registration fee is paid (announcement 1193). Physical and electronic certificates were both valid for importing and placing devices on the market during the transition.
  • 1 November 2023: electronic certificates in full (announcement 1296), ahead of the 1 January 2024 date first announced. MDA no longer issues physical registration certificates.

Where the certificate has to show up

  • On the label and at the point of sale. Within 6 months of registration, the registration number, the manufacturer and authorised representative details and the QR code from the registration certificate must be visible to customers at the point of sale (MDA/GD/0026, 7th Edition, section 4.1). See labelling requirements.
  • In adverts. An advert must carry the statement “Registered under Act 737” and the registration number, plus the approval number where the advert needed MDA approval (MDA/GL/04). See advertisement approval.
  • In tenders. Under MDA Circular Letter No. 1/2026, a healthcare facility buying by tender must obtain the tenderer’s distributor licence, the registration certificate of each device and a letter appointing the tenderer from the authorised representative or distributor. See KKM and hospital tenders.
  • At re-registration. A CAB can recertify by way of verification only while the registration certificate is still valid (MDA/GD/0070, Table 1). See renewals.
  • For changes. MDA charges change notification fees per registration certificate (MDA/GD/0020), and when a device moves to a new authorised representative the current AR surrenders the certificates to MDA (MDA/GD/0041).

The certificate belongs to the registration holder. For a foreign manufacturer that is its authorised representative, not the manufacturer and not each distributor. Distributors work from a copy plus an appointment letter.

The establishment licence

Section 15 of Act 737 bars a company from importing, exporting or placing a registered device on the market without an establishment licence, and MDA grants it under section 18 for a prescribed period: three years from issue, and three years from each renewal (regulations 12(3) and 13(5)). Since 1 July 2024 each role is a separate licence, so a company that imports and distributes holds two.

  • 5 June 2024: MDA began issuing electronic establishment licences, downloadable from MeDC@St 2.0+ once the application is complete and the licence fee is paid; physical licences were still issued for collection at MDA’s office (announcement 1416).
  • 16 April 2026: electronic licences in full. Applications approved before that date also get a physical licence, collected at MDA’s office 7 working days after the application is completed; those approved after 15 April 2026 get only the electronic licence (announcement 1789).
  • Pay before you can download. The electronic licence is released once the licence fee is paid. MDA’s notice of 3 November 2026 says 2026 licence fee payments must be completed by 14 December 2026, because its account closes on 21 December.
  • Personal to the licensee. Section 21 bars assigning, sub-licensing or transferring a licence without MDA’s prior written approval: a fine of RM50,000 to RM500,000, up to 3 years in prison, or both.

How to apply, the fees by role and the GDPMD prerequisite are on our establishment licence service page; our sister site has the full MDA establishment licence guide.

How to check an MDA certificate

MDA’s public register, the Malaysia Medical Device Register (MMDR), has a search tab for each kind of record:

You were givenSearch this MMDR tabMatch
A device registration certificateRegistered Medical DeviceRegistration number, device name, registration holder, validity dates
A supplier’s or distributor’s licenceLicensed EstablishmentCompany name, licence number and the role it covers
A CAB report or GDPMD certificateRegistered CABThat the CAB is registered with MDA
A special access or demonstration exemptionSpecial Access, DemonstrationDevice, applicant and period

Then check MDA’s list of suspended or revoked licences and cancelled devices. A certificate that looks right but is missing from MMDR, or names a different holder, is a reason to stop and ask. Our sister site walks through the search screen by screen in how to check a medical device registration.

MDA approved, registered or certified?

For a device the legal status is registered: section 5 of Act 737 bars placing an unregistered device on the market, and section 7 is where MDA registers it. The wording MDA requires on adverts is “Registered under Act 737”, and the label must state that the device is registered under the Act. Companies are licensed, not registered. “MDA approved” and “MDA certified” are everyday shorthand, not statuses in the Act. In practice:

  • A device with a registration number in MMDR is what people mean by “MDA approved”.
  • A company with a GDPMD or ISO 13485 certificate has passed a quality audit. That does not make its devices registered or the company licensed.
  • A device approved abroad (CE, FDA, HSA) still needs Malaysian registration; foreign approvals can shorten the route through verification, not replace it.

“KKM approved” means the same registration for a device; medicines, supplements and cosmetics are approved by NPRA instead. See KKM approval: which agency approves what.

When a certificate expires, is cancelled or changes

Expiry

Re-registration opens in MeDC@St 2.0+ one year before the certificate expires. The system deactivates an expired certificate automatically, and an expired device can only come back through a new registration with a full conformity assessment. Eligible devices have had an automated re-registration route since 13 July 2026. Establishment licences are renewed for another three years; section 24(1) lets a licensee apply “not later than one year before the expiry date”. Details on our renewals page.

Cancellation

MDA may cancel a registration if the holder breaks the Act or the regulations, breaches a registration condition or is convicted of an offence under the Act (section 9(1)). It must first give written notice and a chance to show cause, and all importation and supply of the device by that establishment is suspended from the day the notice is received (section 9(2)). The holder has 90 days to show cause (regulation 7(2)), and a cancelled holder gets no compensation and no refund of the registration fee (section 9(3)). Licences can be suspended or revoked on the same grounds (section 22).

Changes and new holders

Changes to a registered device are sorted into three categories under MDA/GD/0020: Category 1 needs a new registration, Category 2 needs MDA’s endorsement before you implement it, and Category 3 can be implemented once complete documents are submitted. See change notification. Moving a registration to a new authorised representative goes through change of ownership under MDA/GD/0041 (RM500 per registration), which keeps the original validity.

Lost certificate

Registration certificates issued since November 2023 and licences issued since June 2024 live in MeDC@St 2.0+: download the copy again from the application record. If your AR holds the registration, it is in the AR’s account.

Who asks for which certificate

Who is askingWhat they need
Government or private hospital tenderYour distributor licence, the registration certificate of each device and the AR’s or distributor’s appointment letter (Circular 1/2026)
Clinic, hospital or dealer buying from youRegistration certificate (or the registration number to check in MMDR) and your licence
Shopee, Lazada or TikTok ShopA registered device; MDA has asked platforms to keep unregistered devices off. See selling medical devices online
A regulator abroadCertificate of Free Sale or Export Certificate
Your CAB at re-registrationA registration certificate that is still valid
MDA, for an advert to the publicThe registration number, then an advertisement approval
Jawapan ringkas · Bahasa Malaysia

Apakah sijil MDA dan bagaimana menyemaknya?

Sijil MDA biasanya bermaksud sijil pendaftaran peranti perubatan (seksyen 7(1) Akta 737), sah selama 5 tahun dan dikeluarkan secara elektronik sepenuhnya sejak 1 November 2023 melalui MeDC@St 2.0+, atau lesen establismen syarikat (satu lesen bagi setiap peranan, sah 3 tahun, elektronik sahaja bagi kelulusan selepas 15 April 2026). Semak kedua-duanya di carian awam MMDR (mdar.mda.gov.my). Sijil GDPMD atau ISO 13485 bukan sijil MDA. WhatsApp 010-206 2070.

FAQ

Frequently asked questions

What is an MDA certificate?
Usually one of two documents issued by Malaysia’s Medical Device Authority under the Medical Device Act 2012 (Act 737). The medical device registration certificate is issued for each registered device: section 7(1) says MDA registers the device, assigns it a registration number and issues a certificate of registration. The establishment licence is issued to a company for each role it plays (manufacturer, authorised representative, importer or distributor). MDA also issues Certificates of Free Sale, Manufacturing Certificates and Export Certificates for exports, product classification letters, advertisement approvals and exemption letters. GDPMD and ISO 13485 certificates are not MDA certificates: conformity assessment bodies and certification bodies issue them.
How do I download my MDA registration certificate?
From MeDC@St 2.0+, MDA’s online system, once the application status is complete and the registration fee has been paid. MDA started issuing electronic registration certificates on 2 May 2023 (announcement 1193) and has issued them only electronically since 1 November 2023 (announcement 1296). If you used an authorised representative, the certificate sits in the AR’s MeDC@St account, because the AR is the registration holder; ask the AR for a copy.
How long is an MDA certificate valid?
It depends on the certificate. A device registration is kept on the Register for 5 years from the date of registration unless MDA cancels it (regulation 6(3), Medical Device Regulations 2012). An establishment licence lasts 3 years from issue and 3 years from each renewal (regulations 12(3) and 13(5)). A Certificate of Free Sale or Export Certificate is issued for 2 or 5 years, as you choose when you pay. Product classification letters issued from 1 October 2026 are valid for 5 years. An advertisement approval lapses when the device registration expires. A special access exemption letter lasts 6 months.
How do I check whether an MDA certificate is valid?
Search MDA’s public register, the Malaysia Medical Device Register (MMDR), at mdar.mda.gov.my. Its tabs cover registered medical devices, licensed establishments, registered CABs, special access, custom-made devices, demonstration units, clinical research and education. Match the registration number, the registration holder and the device name with the certificate you were given, and check the validity dates. MDA also publishes a list of suspended or revoked establishment licences and cancelled medical devices (announcement 511).
Is a GDPMD certificate the same as an MDA licence?
No. A GDPMD certificate is issued by a conformity assessment body after it audits your distribution quality system. It is a prerequisite for the establishment licence of an importer, distributor or authorised representative, but it does not let you import or sell devices. Only the MDA establishment licence does that, and every device still needs its own registration certificate.
What happens if my registration certificate expires?
MeDC@St 2.0+ deactivates an expired certificate automatically, and the device cannot be re-registered: it needs a new registration application with a full conformity assessment, and it cannot lawfully be placed on the market in the meantime. Re-registration opens in the system one year before the expiry date, so apply in the final year.
Can an MDA certificate be transferred to another company?
An establishment licence cannot. Section 21 of Act 737 makes it personal to the licensee; assigning, sub-licensing or transferring it without MDA’s prior written approval is an offence carrying a fine of RM50,000 to RM500,000, up to 3 years in prison, or both. A device registration can move to a new registration holder through MDA’s change of ownership process under MDA/GD/0041 (RM500 per registration), without registering the device again.
Does MDA still issue physical certificates?
Not for new device registrations, which have been electronic only since 1 November 2023. Establishment licences approved before 16 April 2026 came with a physical licence, collected at MDA’s office 7 working days after the application was completed; licences approved after 15 April 2026 are electronic only (announcement 1789). Certificates of Free Sale, Manufacturing Certificates and Export Certificates have been issued electronically, by email within 7 working days of payment, since 1 March 2026 (announcement 1746).

Need a certificate, or one is about to expire?

Send us the device or your registration numbers. We tell you which MDA certificate the buyer, tender or platform needs, get it registered or renewed, and keep every expiry date on one calendar.

WhatsApp