MDA certificates: what each one proves, how to get it and how to check it.
“MDA certificate” can mean six different documents. Buyers, hospitals, tender boards, online platforms and foreign regulators each ask for a different one. Here is every certificate the Medical Device Authority issues, how long it lasts, where to download it and how to check one you have been given.
Short answer: an MDA certificate is usually the medical device registration certificate, issued for each device under section 7(1) of Act 737 with its registration number. It is valid for 5 years and has been electronic only since 1 November 2023, downloaded from MeDC@St 2.0+. The other one people mean is the company’s establishment licence (one per role, valid 3 years). Check either in MDA’s public MMDR search. A GDPMD or ISO 13485 certificate is not issued by MDA and does not let you sell a device.
Every certificate MDA issues
Fees for each licence and approval are in our medical device licensing table. This table is about the documents themselves.
| Certificate | What it proves | Who holds it | Valid for | Format | How to check |
|---|---|---|---|---|---|
| Medical device registration certificate | The device is registered under Act 737 and has a registration number (s.7(1)) | The registration holder: the local manufacturer, or the authorised representative of a foreign manufacturer | 5 years from registration (reg. 6(3)) | Electronic only since 1 Nov 2023; download from MeDC@St 2.0+ | MMDR: Registered Medical Device |
| Establishment licence | The company may act in one role: manufacturer, authorised representative, importer or distributor (s.15); one licence per role since 1 Jul 2024 | The company | 3 years from issue and from each renewal (regs. 12(3), 13(5)) | Electronic from 5 Jun 2024; electronic only for approvals after 15 Apr 2026 | MMDR: Licensed Establishment |
| Certificate of Free Sale / Manufacturing Certificate | A registered device is lawfully sold in, or manufactured in, Malaysia (for importing regulators abroad) | The exporter or manufacturer | 2 or 5 years, chosen at payment | Electronic by email since 1 Mar 2026 | Ask MDA’s International and Industry Affairs Division |
| Export Certificate | A device made or imported only for export is exempt under the Exemption Order 2024 | The exporter | 2 or 5 years, chosen at payment | Electronic by email since 1 Mar 2026 | Ask MDA’s International and Industry Affairs Division |
| Product classification letter | Whether a product is a medical device at all (it does not assign a risk class) | The applicant | 5 years for letters issued from 1 Oct 2026 | By email from MDA’s online classification system since 1 Aug 2026 | The letter itself |
| Advertisement approval | An advert for a registered device was approved; the ad carries the approval number | The advertiser | Until the device registration expires | Approval with an approval number | MDA’s published list of advertisement approvals |
| Exemption letters (special access, export only, demonstration, custom-made) | A named device or use is exempt under the Medical Device (Exemption) Order 2024 | The applicant named for each route | By route: special access 6 months, export-only 5 years | Letter from MDA | MMDR: Special Access, Custom Made, Demonstration tabs |
Not MDA certificates: the GDPMD certificate (issued by a conformity assessment body after a distribution audit), the ISO 13485 certificate (issued by a certification body) and a CAB’s conformity assessment report. They are inputs to an MDA licence or registration, not substitutes for one.
The medical device registration certificate
Section 7(1) of the Medical Device Act 2012 (Act 737) says that once a device has been through conformity assessment by a CAB and the fee is paid, MDA may register the device for a prescribed period, assign it a registration number and issue the applicant a certificate of registration. Regulation 6(3) of the Medical Device Regulations 2012 sets the period: the device stays on the Register for five years from the date of registration unless MDA cancels it.
Electronic since 2023
- 2 May 2023: MDA began issuing electronic registration certificates, downloadable from MeDC@St 2.0+ once the application is complete and the registration fee is paid (announcement 1193). Physical and electronic certificates were both valid for importing and placing devices on the market during the transition.
- 1 November 2023: electronic certificates in full (announcement 1296), ahead of the 1 January 2024 date first announced. MDA no longer issues physical registration certificates.
Where the certificate has to show up
- On the label and at the point of sale. Within 6 months of registration, the registration number, the manufacturer and authorised representative details and the QR code from the registration certificate must be visible to customers at the point of sale (MDA/GD/0026, 7th Edition, section 4.1). See labelling requirements.
- In adverts. An advert must carry the statement “Registered under Act 737” and the registration number, plus the approval number where the advert needed MDA approval (MDA/GL/04). See advertisement approval.
- In tenders. Under MDA Circular Letter No. 1/2026, a healthcare facility buying by tender must obtain the tenderer’s distributor licence, the registration certificate of each device and a letter appointing the tenderer from the authorised representative or distributor. See KKM and hospital tenders.
- At re-registration. A CAB can recertify by way of verification only while the registration certificate is still valid (MDA/GD/0070, Table 1). See renewals.
- For changes. MDA charges change notification fees per registration certificate (MDA/GD/0020), and when a device moves to a new authorised representative the current AR surrenders the certificates to MDA (MDA/GD/0041).
The certificate belongs to the registration holder. For a foreign manufacturer that is its authorised representative, not the manufacturer and not each distributor. Distributors work from a copy plus an appointment letter.
The establishment licence
Section 15 of Act 737 bars a company from importing, exporting or placing a registered device on the market without an establishment licence, and MDA grants it under section 18 for a prescribed period: three years from issue, and three years from each renewal (regulations 12(3) and 13(5)). Since 1 July 2024 each role is a separate licence, so a company that imports and distributes holds two.
- 5 June 2024: MDA began issuing electronic establishment licences, downloadable from MeDC@St 2.0+ once the application is complete and the licence fee is paid; physical licences were still issued for collection at MDA’s office (announcement 1416).
- 16 April 2026: electronic licences in full. Applications approved before that date also get a physical licence, collected at MDA’s office 7 working days after the application is completed; those approved after 15 April 2026 get only the electronic licence (announcement 1789).
- Pay before you can download. The electronic licence is released once the licence fee is paid. MDA’s notice of 3 November 2026 says 2026 licence fee payments must be completed by 14 December 2026, because its account closes on 21 December.
- Personal to the licensee. Section 21 bars assigning, sub-licensing or transferring a licence without MDA’s prior written approval: a fine of RM50,000 to RM500,000, up to 3 years in prison, or both.
How to apply, the fees by role and the GDPMD prerequisite are on our establishment licence service page; our sister site has the full MDA establishment licence guide.
How to check an MDA certificate
MDA’s public register, the Malaysia Medical Device Register (MMDR), has a search tab for each kind of record:
| You were given | Search this MMDR tab | Match |
|---|---|---|
| A device registration certificate | Registered Medical Device | Registration number, device name, registration holder, validity dates |
| A supplier’s or distributor’s licence | Licensed Establishment | Company name, licence number and the role it covers |
| A CAB report or GDPMD certificate | Registered CAB | That the CAB is registered with MDA |
| A special access or demonstration exemption | Special Access, Demonstration | Device, applicant and period |
Then check MDA’s list of suspended or revoked licences and cancelled devices. A certificate that looks right but is missing from MMDR, or names a different holder, is a reason to stop and ask. Our sister site walks through the search screen by screen in how to check a medical device registration.
MDA approved, registered or certified?
For a device the legal status is registered: section 5 of Act 737 bars placing an unregistered device on the market, and section 7 is where MDA registers it. The wording MDA requires on adverts is “Registered under Act 737”, and the label must state that the device is registered under the Act. Companies are licensed, not registered. “MDA approved” and “MDA certified” are everyday shorthand, not statuses in the Act. In practice:
- A device with a registration number in MMDR is what people mean by “MDA approved”.
- A company with a GDPMD or ISO 13485 certificate has passed a quality audit. That does not make its devices registered or the company licensed.
- A device approved abroad (CE, FDA, HSA) still needs Malaysian registration; foreign approvals can shorten the route through verification, not replace it.
“KKM approved” means the same registration for a device; medicines, supplements and cosmetics are approved by NPRA instead. See KKM approval: which agency approves what.
When a certificate expires, is cancelled or changes
Expiry
Re-registration opens in MeDC@St 2.0+ one year before the certificate expires. The system deactivates an expired certificate automatically, and an expired device can only come back through a new registration with a full conformity assessment. Eligible devices have had an automated re-registration route since 13 July 2026. Establishment licences are renewed for another three years; section 24(1) lets a licensee apply “not later than one year before the expiry date”. Details on our renewals page.
Cancellation
MDA may cancel a registration if the holder breaks the Act or the regulations, breaches a registration condition or is convicted of an offence under the Act (section 9(1)). It must first give written notice and a chance to show cause, and all importation and supply of the device by that establishment is suspended from the day the notice is received (section 9(2)). The holder has 90 days to show cause (regulation 7(2)), and a cancelled holder gets no compensation and no refund of the registration fee (section 9(3)). Licences can be suspended or revoked on the same grounds (section 22).
Changes and new holders
Changes to a registered device are sorted into three categories under MDA/GD/0020: Category 1 needs a new registration, Category 2 needs MDA’s endorsement before you implement it, and Category 3 can be implemented once complete documents are submitted. See change notification. Moving a registration to a new authorised representative goes through change of ownership under MDA/GD/0041 (RM500 per registration), which keeps the original validity.
Lost certificate
Registration certificates issued since November 2023 and licences issued since June 2024 live in MeDC@St 2.0+: download the copy again from the application record. If your AR holds the registration, it is in the AR’s account.
Who asks for which certificate
| Who is asking | What they need |
|---|---|
| Government or private hospital tender | Your distributor licence, the registration certificate of each device and the AR’s or distributor’s appointment letter (Circular 1/2026) |
| Clinic, hospital or dealer buying from you | Registration certificate (or the registration number to check in MMDR) and your licence |
| Shopee, Lazada or TikTok Shop | A registered device; MDA has asked platforms to keep unregistered devices off. See selling medical devices online |
| A regulator abroad | Certificate of Free Sale or Export Certificate |
| Your CAB at re-registration | A registration certificate that is still valid |
| MDA, for an advert to the public | The registration number, then an advertisement approval |
Apakah sijil MDA dan bagaimana menyemaknya?
Sijil MDA biasanya bermaksud sijil pendaftaran peranti perubatan (seksyen 7(1) Akta 737), sah selama 5 tahun dan dikeluarkan secara elektronik sepenuhnya sejak 1 November 2023 melalui MeDC@St 2.0+, atau lesen establismen syarikat (satu lesen bagi setiap peranan, sah 3 tahun, elektronik sahaja bagi kelulusan selepas 15 April 2026). Semak kedua-duanya di carian awam MMDR (mdar.mda.gov.my). Sijil GDPMD atau ISO 13485 bukan sijil MDA. WhatsApp 010-206 2070.
Frequently asked questions
What is an MDA certificate?
How do I download my MDA registration certificate?
How long is an MDA certificate valid?
How do I check whether an MDA certificate is valid?
Is a GDPMD certificate the same as an MDA licence?
What happens if my registration certificate expires?
Can an MDA certificate be transferred to another company?
Does MDA still issue physical certificates?
Need a certificate, or one is about to expire?
Send us the device or your registration numbers. We tell you which MDA certificate the buyer, tender or platform needs, get it registered or renewed, and keep every expiry date on one calendar.