If any part of your device touches a patient or user, MDA expects biocompatibility evidence in the CSDT. The governing standard, ISO 10993-1, was rewritten in November 2025. Here is what MDA asks for and what the new edition changes.

Short answer: the Medical Device Regulations 2012 require “biocompatibility tests conducted on materials used in a medical device” in the CSDT, and MDA/GD/0070 lists them for Class B, C and D. MDA/GD/0008 says the tests must at minimum be on the finished, sterilised device. MDA does not name an ISO 10993 edition. The current one is ISO 10993-1:2025 (18 November 2025), which treats biological evaluation as part of ISO 14971 risk management. If your evaluation uses an older edition, MDA/GD/0070 asks for a gap assessment.

What MDA asks for

  • Essential Principle 8 (Medical Device Regulations 2012, Third Schedule, Appendix 1): design and manufacture must pay particular attention to the choice of materials, especially toxicity, and the compatibility between materials and biological tissues, cells and body fluids. Paragraph 8(2) adds contaminants and residues, the duration and frequency of exposure, and substances that may leach from the device.
  • The CSDT (Third Schedule, Appendix 2, paragraph 4(2)): pre-clinical documentation “shall include the report and/or certification and/or declaration of” biocompatibility tests on materials used in the device.
  • MDA/GD/0008 (CSDT guidance, First Edition, March 2014): tests on samples from the finished, sterilised device; materials that are significantly different must be characterised; the device description must give a complete chemical, biological and physical characterisation of materials in direct or indirect body contact; pre-clinical animal studies must follow good laboratory practices.
  • MDA/GD/0070 (Second Edition, October 2025), Annex 4, for Class B, C and D: the pre-clinical section shall include biocompatibility tests on materials, physical tests, animal studies where needed and software verification. Where outdated standards were applied, a gap assessment to show the state of the art.

Some MDA device guidance names ISO 10993 directly: MDA/GD/0033 for face masks says to complete a biocompatibility evaluation according to ISO 10993, and MDA/GD/0058 for PPE cites ISO 10993-10 and ISO 10993-23 for face shields. See our face mask and PPE guide.

Biological evaluation is a risk process, not a test list

ISO 10993-1:2025 says plainly that biological evaluation “is an exercise in risk management”, carried out within the ISO 14971 process. Its clauses run:

  1. Biological evaluation plan (clause 5).
  2. Biological risk analysis (clause 6): material and chemical characterisation, the categorisation of body contact and exposure, the biological effects to address, a gap analysis of existing data, biological equivalence to other devices, and testing only where gaps remain.
  3. Biological risk evaluation and control (clauses 7 and 8).
  4. Biological evaluation report (clause 9).
  5. Production and post-production information (clause 10).

In practice this means the file starts with chemistry and existing data, and animal or in vitro tests fill the gaps. The 2025 edition also says it does not mandate re-testing devices already on the market with established and acceptable safety profiles. Our ISO 14971 guide covers the risk framework it sits in.

How contact and duration are categorised

The familiar scheme from the 2018 edition, as reproduced in FDA’s 2023 guidance:

Nature of body contactContact
Surface deviceIntact skin; mucosal membrane; breached or compromised surface
Externally communicating deviceBlood path, indirect; tissue, bone or dentin; circulating blood
Implant deviceTissue or bone; blood

Duration: limited (up to 24 hours), prolonged (more than 24 hours to 30 days) and long term (more than 30 days). The combination decides which biological effects to address, from cytotoxicity, sensitisation and irritation for every contacting device, up to systemic toxicity, genotoxicity, implantation effects, haemocompatibility, chronic toxicity and carcinogenicity for longer and deeper contact.

What the 2025 edition changed

  • The externally communicating category is gone. Contact is described by tissue type: intact skin, intact mucosa, breached or compromised surfaces and internal tissues other than blood, and circulating blood.
  • Exposure duration now accounts for daily, intermittent and cumulative use, with new defined terms for a contact day and the total exposure period.
  • The biological effects (previously “end points”) were revised, with a new Annex B explaining why. FDA’s summary: genotoxicity is now expected for all devices with prolonged contact except intact-skin-only devices, carcinogenicity for long-term intact mucosal contact, and material-mediated pyrogenicity moved out of the main table to “other biological effects”.
  • “Effects after implantation” became “local effects after tissue contact”.
  • Annex A now covers materials characterisation only.

Which edition should your file use?

AuthorityPosition on ISO 10993-1 (checked 11 November 2026)
MDA (Malaysia)No edition named; MDA/GD/0070 asks for a gap assessment where outdated standards were applied
US FDA2025 edition recognised in part from 25 May 2026; declarations to the 2018 edition accepted until 1 July 2029
EU (MDR)EN ISO 10993-1:2025 harmonised from 17 June 2026

Our reading: a new evaluation should follow the 2025 edition. An existing evaluation to the 2018 edition is still the normal case in 2026 submissions; include a short gap assessment against 2025 rather than re-testing by default.

Current editions of the parts you will cite

PartCoversCurrent edition
ISO 10993-1Requirements and general principles; the evaluation framework2025 (Edition 6, 18 November 2025)
ISO 10993-3Genotoxicity, carcinogenicity, reproductive and developmental toxicity2026 (Edition 4, August 2026)
ISO 10993-4Interactions with blood (haemocompatibility)2017 + Amd 1:2025; a revision is in development
ISO 10993-5In vitro cytotoxicity2009 (confirmed 2022)
ISO 10993-7Ethylene oxide sterilisation residuals2026 (Edition 3, April 2026)
ISO 10993-10Skin sensitisation2021 (Edition 4)
ISO 10993-11Systemic toxicity, including material-mediated pyrogenicity2017; a revision is in development
ISO 10993-12Sample preparation and reference materials2021 + Amd 1:2025
ISO 10993-17Toxicological risk assessment of constituents2023 + Amd 1:2025
ISO 10993-18Chemical characterisation of materials2020 + Amd 1:2022 (confirmed 2025)
ISO 10993-23Irritation2021 + Amd 1:2025 (confirmed 2026)

From ISO’s standard pages, 11 November 2026. The Essential Principles checklist should give the date of each standard you cite (Third Schedule, Table 1 note; MDA/GD/0008). See our Essential Principles checklist guide.

After registration

  • Material changes. MDA’s change management guidance, MDA/GD/0072 (First Edition, May 2026, not yet in force), treats a change of material or material formulation in a device that contacts body tissues or fluids as significant, with pre-clinical reports and an updated risk management report as evidence. See change notification.
  • Sterilisation. Ethylene oxide residuals are assessed under ISO 10993-7, which was revised in April 2026. The validation itself is covered in our sterilisation validation guide.

Our role

We plan the biological evaluation, review lab reports and write the evaluation into the CSDT as part of our registration fee (Class B RM6,800, Class C RM9,800, Class D RM12,000, with CAB and MDA fees). Lab testing is paid to the lab. For a standalone gap assessment, send your reports on WhatsApp for a quote. See our CSDT guide.

Jawapan ringkas · Bahasa Malaysia

Adakah MDA memerlukan ujian biokeserasian ISO 10993?

Ya, bagi peranti yang bahannya bersentuhan dengan badan. Peraturan-Peraturan Peranti Perubatan 2012 memasukkan ujian biokeserasian bahan dalam bahagian pra-klinikal CSDT, dan MDA/GD/0008 menetapkan ujian dibuat sekurang-kurangnya pada sampel peranti siap yang telah disteril. MDA tidak menamakan edisi ISO 10993. Edisi semasa ialah ISO 10993-1:2025 (18 November 2025). Jika edisi lama digunakan, MDA/GD/0070 meminta penilaian jurang (gap assessment). WhatsApp 010-206 2070.

FAQ

Frequently asked questions

Does MDA require ISO 10993 biocompatibility testing?
MDA requires biocompatibility evidence for devices whose materials contact the body. The Medical Device Regulations 2012 put “biocompatibility tests conducted on materials used in a medical device” in the CSDT’s pre-clinical section, and MDA/GD/0070 lists them for Class B, C and D. MDA does not name an ISO 10993 edition, but ISO 10993 is the standard used worldwide for this, and MDA’s own device guidance for face masks says to complete an evaluation according to ISO 10993.
What is the current edition of ISO 10993-1?
ISO 10993-1:2025, the sixth edition, published on 18 November 2025. It replaced ISO 10993-1:2018 and is retitled “Requirements and general principles for the evaluation of biological safety within a risk management process”. It reorganises the evaluation around ISO 14971 and changes how body contact and biological effects are categorised.
Can I still use a biological evaluation done to ISO 10993-1:2018 for MDA?
MDA has not published a transition rule for ISO 10993-1. Its registration submission guide MDA/GD/0070 says that if outdated standards were applied, a gap assessment needs to be provided to demonstrate the state of the art. For comparison, the US FDA will accept declarations of conformity to the 2018 edition until 1 July 2029, and the EU listed EN ISO 10993-1:2025 as harmonised from 17 June 2026. ISO 10993-1:2025 itself says it does not mandate re-testing of devices already on the market with established and acceptable safety profiles.
Do biocompatibility tests have to be on the finished device?
Yes, as a minimum. MDA’s CSDT guidance MDA/GD/0008 says biocompatibility tests must be conducted on samples from the finished, sterilised device, and that all materials that are significantly different must be characterised. Testing raw materials alone is not enough.
Do Class A devices need biocompatibility evidence?
Class A applications do not upload a CSDT, and MDA’s Class A requirements mention biological evaluation tests only as an optional attachment where applicable. But the Essential Principles on materials and biocompatibility apply to every class, the manufacturer must keep a summary of technical documentation available on request, and the Declaration of Conformity covers all the Essential Principles. A body-contacting Class A device still needs its biological evaluation on file.
Does MDA require a GLP or ISO 17025 accredited laboratory?
MDA/GD/0008 says pre-clinical animal studies must be undertaken using good laboratory practices. We found no MDA requirement naming ISO/IEC 17025 accreditation for biocompatibility labs. Use a lab whose reports are accepted in your reference markets, because the same report usually serves the EU, US and Malaysian files.
Is there a Malaysian Standard for ISO 10993?
Several parts, all behind the ISO editions. The Department of Standards Malaysia catalogue lists MS ISO 10993-1:2011, an adoption of ISO 10993-1:2009 confirmed in 2015, and MS adoptions of parts 4, 10, 14, 15, 16 and 17 based on 2000–2010 ISO editions. The MS versions of parts 5, 9 and 13 are withdrawn. Work to the current ISO editions.

Biocompatibility that a CAB can follow

We turn your materials data and lab reports into a biological evaluation that maps to MDA’s CSDT and the current ISO 10993-1.