ISO 10993 biocompatibility for MDA registration
If any part of your device touches a patient or user, MDA expects biocompatibility evidence in the CSDT. The governing standard, ISO 10993-1, was rewritten in November 2025. Here is what MDA asks for and what the new edition changes.
Short answer: the Medical Device Regulations 2012 require “biocompatibility tests conducted on materials used in a medical device” in the CSDT, and MDA/GD/0070 lists them for Class B, C and D. MDA/GD/0008 says the tests must at minimum be on the finished, sterilised device. MDA does not name an ISO 10993 edition. The current one is ISO 10993-1:2025 (18 November 2025), which treats biological evaluation as part of ISO 14971 risk management. If your evaluation uses an older edition, MDA/GD/0070 asks for a gap assessment.
What MDA asks for
- Essential Principle 8 (Medical Device Regulations 2012, Third Schedule, Appendix 1): design and manufacture must pay particular attention to the choice of materials, especially toxicity, and the compatibility between materials and biological tissues, cells and body fluids. Paragraph 8(2) adds contaminants and residues, the duration and frequency of exposure, and substances that may leach from the device.
- The CSDT (Third Schedule, Appendix 2, paragraph 4(2)): pre-clinical documentation “shall include the report and/or certification and/or declaration of” biocompatibility tests on materials used in the device.
- MDA/GD/0008 (CSDT guidance, First Edition, March 2014): tests on samples from the finished, sterilised device; materials that are significantly different must be characterised; the device description must give a complete chemical, biological and physical characterisation of materials in direct or indirect body contact; pre-clinical animal studies must follow good laboratory practices.
- MDA/GD/0070 (Second Edition, October 2025), Annex 4, for Class B, C and D: the pre-clinical section shall include biocompatibility tests on materials, physical tests, animal studies where needed and software verification. Where outdated standards were applied, a gap assessment to show the state of the art.
Some MDA device guidance names ISO 10993 directly: MDA/GD/0033 for face masks says to complete a biocompatibility evaluation according to ISO 10993, and MDA/GD/0058 for PPE cites ISO 10993-10 and ISO 10993-23 for face shields. See our face mask and PPE guide.
Biological evaluation is a risk process, not a test list
ISO 10993-1:2025 says plainly that biological evaluation “is an exercise in risk management”, carried out within the ISO 14971 process. Its clauses run:
- Biological evaluation plan (clause 5).
- Biological risk analysis (clause 6): material and chemical characterisation, the categorisation of body contact and exposure, the biological effects to address, a gap analysis of existing data, biological equivalence to other devices, and testing only where gaps remain.
- Biological risk evaluation and control (clauses 7 and 8).
- Biological evaluation report (clause 9).
- Production and post-production information (clause 10).
In practice this means the file starts with chemistry and existing data, and animal or in vitro tests fill the gaps. The 2025 edition also says it does not mandate re-testing devices already on the market with established and acceptable safety profiles. Our ISO 14971 guide covers the risk framework it sits in.
How contact and duration are categorised
The familiar scheme from the 2018 edition, as reproduced in FDA’s 2023 guidance:
| Nature of body contact | Contact |
|---|---|
| Surface device | Intact skin; mucosal membrane; breached or compromised surface |
| Externally communicating device | Blood path, indirect; tissue, bone or dentin; circulating blood |
| Implant device | Tissue or bone; blood |
Duration: limited (up to 24 hours), prolonged (more than 24 hours to 30 days) and long term (more than 30 days). The combination decides which biological effects to address, from cytotoxicity, sensitisation and irritation for every contacting device, up to systemic toxicity, genotoxicity, implantation effects, haemocompatibility, chronic toxicity and carcinogenicity for longer and deeper contact.
What the 2025 edition changed
- The externally communicating category is gone. Contact is described by tissue type: intact skin, intact mucosa, breached or compromised surfaces and internal tissues other than blood, and circulating blood.
- Exposure duration now accounts for daily, intermittent and cumulative use, with new defined terms for a contact day and the total exposure period.
- The biological effects (previously “end points”) were revised, with a new Annex B explaining why. FDA’s summary: genotoxicity is now expected for all devices with prolonged contact except intact-skin-only devices, carcinogenicity for long-term intact mucosal contact, and material-mediated pyrogenicity moved out of the main table to “other biological effects”.
- “Effects after implantation” became “local effects after tissue contact”.
- Annex A now covers materials characterisation only.
Which edition should your file use?
| Authority | Position on ISO 10993-1 (checked 11 November 2026) |
|---|---|
| MDA (Malaysia) | No edition named; MDA/GD/0070 asks for a gap assessment where outdated standards were applied |
| US FDA | 2025 edition recognised in part from 25 May 2026; declarations to the 2018 edition accepted until 1 July 2029 |
| EU (MDR) | EN ISO 10993-1:2025 harmonised from 17 June 2026 |
Our reading: a new evaluation should follow the 2025 edition. An existing evaluation to the 2018 edition is still the normal case in 2026 submissions; include a short gap assessment against 2025 rather than re-testing by default.
Current editions of the parts you will cite
| Part | Covers | Current edition |
|---|---|---|
| ISO 10993-1 | Requirements and general principles; the evaluation framework | 2025 (Edition 6, 18 November 2025) |
| ISO 10993-3 | Genotoxicity, carcinogenicity, reproductive and developmental toxicity | 2026 (Edition 4, August 2026) |
| ISO 10993-4 | Interactions with blood (haemocompatibility) | 2017 + Amd 1:2025; a revision is in development |
| ISO 10993-5 | In vitro cytotoxicity | 2009 (confirmed 2022) |
| ISO 10993-7 | Ethylene oxide sterilisation residuals | 2026 (Edition 3, April 2026) |
| ISO 10993-10 | Skin sensitisation | 2021 (Edition 4) |
| ISO 10993-11 | Systemic toxicity, including material-mediated pyrogenicity | 2017; a revision is in development |
| ISO 10993-12 | Sample preparation and reference materials | 2021 + Amd 1:2025 |
| ISO 10993-17 | Toxicological risk assessment of constituents | 2023 + Amd 1:2025 |
| ISO 10993-18 | Chemical characterisation of materials | 2020 + Amd 1:2022 (confirmed 2025) |
| ISO 10993-23 | Irritation | 2021 + Amd 1:2025 (confirmed 2026) |
From ISO’s standard pages, 11 November 2026. The Essential Principles checklist should give the date of each standard you cite (Third Schedule, Table 1 note; MDA/GD/0008). See our Essential Principles checklist guide.
After registration
- Material changes. MDA’s change management guidance, MDA/GD/0072 (First Edition, May 2026, not yet in force), treats a change of material or material formulation in a device that contacts body tissues or fluids as significant, with pre-clinical reports and an updated risk management report as evidence. See change notification.
- Sterilisation. Ethylene oxide residuals are assessed under ISO 10993-7, which was revised in April 2026. The validation itself is covered in our sterilisation validation guide.
Our role
We plan the biological evaluation, review lab reports and write the evaluation into the CSDT as part of our registration fee (Class B RM6,800, Class C RM9,800, Class D RM12,000, with CAB and MDA fees). Lab testing is paid to the lab. For a standalone gap assessment, send your reports on WhatsApp for a quote. See our CSDT guide.
Adakah MDA memerlukan ujian biokeserasian ISO 10993?
Ya, bagi peranti yang bahannya bersentuhan dengan badan. Peraturan-Peraturan Peranti Perubatan 2012 memasukkan ujian biokeserasian bahan dalam bahagian pra-klinikal CSDT, dan MDA/GD/0008 menetapkan ujian dibuat sekurang-kurangnya pada sampel peranti siap yang telah disteril. MDA tidak menamakan edisi ISO 10993. Edisi semasa ialah ISO 10993-1:2025 (18 November 2025). Jika edisi lama digunakan, MDA/GD/0070 meminta penilaian jurang (gap assessment). WhatsApp 010-206 2070.
Frequently asked questions
Does MDA require ISO 10993 biocompatibility testing?
What is the current edition of ISO 10993-1?
Can I still use a biological evaluation done to ISO 10993-1:2018 for MDA?
Do biocompatibility tests have to be on the finished device?
Do Class A devices need biocompatibility evidence?
Does MDA require a GLP or ISO 17025 accredited laboratory?
Is there a Malaysian Standard for ISO 10993?
Biocompatibility that a CAB can follow
We turn your materials data and lab reports into a biological evaluation that maps to MDA’s CSDT and the current ISO 10993-1.