IEC 62304 for medical device software in Malaysia

Whether it is firmware in a monitor or a diagnostic app, software in a medical device has to come with evidence that its design and development process was controlled. MDA’s rules ask for that evidence; IEC 62304 is the standard for producing it. Here is what MDA requires, how the software safety classes work, what to do about SOUP and legacy code, and what goes into the software file.

Short answer: MDA does not name a software standard, but the Medical Device Regulations 2012 require the CSDT to contain software validation studies: objective evidence that validates the software design and development process, and the results of all verification, validation and testing done in-house and in a user’s environment before release, for every hardware configuration in the labelling. IEC 62304:2006 + A1:2015 (Edition 1.1) is the international standard for that process, and the usual way to produce the evidence. It sorts software into safety classes A, B and C by the harm it could contribute to, and the class decides how much design and testing documentation you need. Edition 2 is still in development.

What MDA asks for on software

WhereWhat it says
Medical Device Regulations 2012, Third Schedule, Appendix 1 (Essential Principles), paragraph 13A device incorporating an electronic programmable system, including software, must be designed to ensure the repeatability, reliability and performance of those systems according to the intended use, with means to eliminate or reduce the risks of a single fault condition
Third Schedule, Appendix 2 (CSDT), paragraph 5The CSDT must contain software validation studies: objective evidence that validates the software design and development process, and the results of all verification, validation and testing in-house and in a user’s environment before final release, for all hardware configurations in the labelling, with representative data from both
MDA/GD/0008 (CSDT guidance), 6.4.2The same evidence, needed “to ensure the correctness of a software product which cannot be fully verified in a finished product”
MDA/GD/0070 (Second Edition, October 2025), Annex 4 (Class B, C and D)Pre-clinical studies include software verification and validation; for IVDs, software verification and validation studies are among the characteristics to validate
MDA/GD/0070, Annex 3 (Class A), section 5Where applicable, other tests for safety or effectiveness, with the purpose, the risk addressed, the methods and the results; software validation is one of the examples given
MDA/GD/0031 (full conformity assessment), CAB checklist item 9Documentation on software validation studies, with the same two parts

We checked MDA/GD/0008, MDA/GD/0031, MDA/GD/0070 and MDA/GD/0072 on 20 November 2026: none names IEC 62304 or any other software standard. The Regulations allow evidence based on standards recognised by MDA or on other standards and the manufacturer’s own tests, and MDA publishes no list of recognised standards. For Class A the full CSDT is not uploaded, but the software validation evidence goes in where applicable and the technical documentation must be available to MDA on request.

What IEC 62304 covers, and what it does not

  • Scope. The life cycle requirements for medical device software, for software that is itself a medical device and software embedded in one. It was prepared jointly by IEC and ISO and published as a double-logo standard.
  • Not covered: validation and final release of the medical device, even when the device consists entirely of software. Those system-level activities sit in product standards such as IEC 60601-1 for electrical equipment and IEC 82304-1 for standalone health software.
  • Assumed foundations: a quality management system and a risk management system. The risk process is ISO 14971, the standard’s only normative reference; IEC 62304 adds the software-specific parts.

The standard is organised into five processes:

ClauseProcessWhat it produces
5Software developmentPlan, requirements, architecture, detailed design, units, integration and system testing, release
6Software maintenanceMaintenance plan, problem and modification analysis, controlled modification and re-release
7Software risk managementSoftware contributions to hazardous situations, risk control measures in software and their verification, risk review of changes
8Software configuration managementIdentified configuration items including SOUP, change control, configuration status records
9Software problem resolutionProblem reports, investigation, trend analysis and verified fixes

The standard treats maintenance as being as important as development, because many field incidents come from inadequate updates and upgrades. Maintenance, configuration management and problem resolution apply to every safety class.

Software safety classes A, B and C

Amendment 1 (2015) made the classification risk-based. The manufacturer assigns each software system a class by the risk of harm from a hazardous situation the software can contribute to in a worst case:

  • Class A: the software cannot contribute to a hazardous situation, or it can, but the risk is not unacceptable once risk control measures external to the software are considered.
  • Class B: it can contribute to a hazardous situation that results in unacceptable risk after external risk controls, and the possible harm is non-serious injury.
  • Class C: as for B, but the possible harm is death or serious injury.

External risk controls can be hardware, an independent software system, healthcare procedures or other means. If you add them, including by changing the system architecture, the software can be reclassified. Where software items are grouped, the group takes the highest class in it unless the risk management file documents a rationale for a lower one.

The software safety class is not the MDA device class. MDA classifies the device (A to D) under the First Schedule rules; see classification. The two are decided separately, so they need not match.

What each class has to do in the development process

ActivityClauseClass AClass BClass C
Software development planning5.1YesYesYes
Software requirements analysis5.2YesYesYes
Software architectural design5.3—YesYes, plus segregation needed for risk control (5.3.5)
Detailed design: dividing the software into units5.4.1—YesYes
Detailed design of each unit and its interfaces5.4.2–5.4.4——Yes
Implementing each software unit5.5.1YesYesYes
Unit verification process and acceptance criteria5.5.2–5.5.5—YesYes, plus additional acceptance criteria (5.5.4)
Software integration and integration testing5.6—YesYes
Software system testing, with traceability to requirements and test records5.7YesYesYes
Software release: verification complete, residual anomalies documented, versions archived5.8YesYesYes

Summary by activity from IEC 62304:2006 + A1:2015. Within an activity some tasks apply only to B and C or only to C (for example, the procedure for identifying and avoiding common software defects, 5.1.12, is B and C). Amendment 1 extended software system testing, documenting known residual anomalies and archiving the released software to Class A, so even Class A software needs documented requirements, system test records and an archived release.

SOUP: third-party and open-source software

SOUP (software of unknown provenance) is software already developed and generally available that was not developed for your device, or software previously developed without adequate records. Operating systems, open-source libraries, communication stacks, SDKs and machine-learning frameworks usually qualify. Your own medical device software system as a whole cannot be called SOUP.

  • All classes: identify each SOUP item by title, manufacturer and a unique designator such as its version (configuration management, 8.1.2).
  • Class B and C: specify the functional and performance requirements each SOUP item must meet, and the hardware and software it needs to run (5.3.3, 5.3.4).
  • Risk: consider SOUP failures and its published anomaly lists when you analyse how the software could contribute to hazardous situations (clause 7).

Legacy software already on the market

Many devices registered in Malaysia run software written before the manufacturer adopted IEC 62304. Amendment 1 added a legacy-software route (4.4) as an alternative to clauses 5 to 9:

  1. Assess post-market feedback, incidents and near incidents, and the risks of continued use.
  2. Run a gap analysis of the available records against the requirements (5.2), architecture (5.3), system testing (5.7) and software risk management (clause 7) for the software’s class.
  3. Decide which missing deliverables would reduce risk, and produce them. The minimum is software system test records.
  4. Document the legacy software version and the rationale for its continued use; handle problems and changes through clauses 9 and 6 from then on.

What goes into the MDA software file

The CSDT carries a summary or the reports themselves; the full IEC 62304 records stay in your design history and are available to the CAB and MDA. A file that will pass review usually holds:

  • Software description: function, version, platform, the hardware configurations in the labelling
  • The software safety class and its rationale, tied to the ISO 14971 risk management file
  • Software development plan and the standards, methods and tools used
  • Software requirements, including risk control measures, user-interface, network and security requirements
  • Architecture and, for Class C, detailed design (Class B and C)
  • Verification records: unit and integration testing (Class B and C), system testing with a traceability matrix from requirements to tests (all classes)
  • Validation in the user’s environment for each hardware configuration, as the Regulations require
  • SOUP list with versions, and the list of known residual anomalies at release
  • Cybersecurity risk assessment for connected software; IEC 81001-5-1 sets out the security activities
  • Release record and the maintenance plan that will govern updates

For standalone software, add the qualification and classification reasoning from our SaMD guide and the clinical evaluation for any diagnostic or therapeutic algorithm. The full dossier structure is on our CSDT technical file page.

Related standards

StandardRoleStatus
IEC 62304:2006 + A1:2015Software life cycle processesEdition 1.1, published 26 June 2015. New edition under development (committee draft stage); stability date 2028
ISO 14971:2019Risk management; normative reference of IEC 62304Current; confirmed by ISO in 2025
IEC 82304-1:2016Product safety of standalone health software, including validation and release at product levelEdition 1.0, published 27 October 2016; revision under development
IEC 81001-5-1:2021Security activities in the health software life cycleEdition 1.0, published 16 December 2021; revision under development (committee draft stage)
IEC 62366-1:2015 + A1:2020Usability engineering, including user-interface requirementsCurrent
IEC TR 80002-1:2009Guidance on applying ISO 14971 to medical device softwareTechnical report, referred to by IEC 62304 for categories of software defects

IEC 62304, IEC 82304-1 and IEC 81001-5-1 checked on the IEC webstore on 20 November 2026; ISO 14971 and IEC 62366-1 as checked for our ISO 14971 and IEC 60601 pages on 11 November 2026. A device with software in a powered unit also needs the electrical safety and EMC evidence on our IEC 60601 page. MDA’s own training arm, now MDA Academy, ran a one-day course on the fundamentals of SaMD at its Cyberjaya office on 15 September 2026 (RM1,000 per participant), covering software validation and testing among other topics.

How we help

  • IEC 62304 gap assessment of your existing software documentation against the standard and MDA’s CSDT requirements, with a list of what to write or test
  • Software safety classification worked out with your engineers from the risk analysis
  • Templates and drafting for the development plan, requirements, architecture, verification records, traceability matrix, SOUP list and release record, fitted to how your team already works
  • Legacy software route for software already on the market elsewhere
  • The software section of the CSDT and the link to the ISO 14971 risk file, then the MDA submission

Preparing the software section is part of our registration fee when we register the device: Class A RM3,800, Class B RM6,800, Class C RM9,800, Class D RM12,000, all-in with CAB and MDA fees. A standalone IEC 62304 gap assessment or documentation build depends on the size and class of the software, so we quote it after seeing what you have. Software companies that will be the legal manufacturer also need an ISO 13485 system (our fee RM7,500 in year 1). See our MDA registration service.

Jawapan ringkas · Bahasa Malaysia

Adakah MDA memerlukan IEC 62304 untuk perisian peranti perubatan?

MDA tidak menamakan piawaian perisian, tetapi Peraturan-Peraturan Peranti Perubatan 2012 (Jadual Ketiga, Lampiran 2, perenggan 5) mewajibkan CSDT mengandungi kajian pengesahan perisian: bukti objektif yang mengesahkan proses reka bentuk dan pembangunan perisian, serta keputusan semua ujian di dalam syarikat dan dalam persekitaran pengguna. IEC 62304:2006 + A1:2015 ialah piawaian antarabangsa untuk proses ini dan cara biasa menghasilkan bukti tersebut. Perisian dikelaskan A, B atau C mengikut kemudaratan yang boleh disumbangkan. WhatsApp 010-206 2070.

FAQ

Frequently asked questions

Does MDA require IEC 62304?
MDA does not name it. None of MDA/GD/0008, MDA/GD/0031, MDA/GD/0070 or MDA/GD/0072 mentions a software standard. What the law requires is evidence: the Medical Device Regulations 2012 (Third Schedule, Appendix 2, paragraph 5) make the CSDT contain software validation studies, including objective evidence that validates the software design and development process. IEC 62304 is the international standard for that process, so it is the usual way to produce the evidence, and a file built to it for other markets carries over.
What is the current edition of IEC 62304? Is Edition 2 out?
The current version is IEC 62304:2006 + Amendment 1:2015, sold as the consolidated Edition 1.1 (published 26 June 2015). IEC’s webstore shows a new edition under development at the committee draft stage, and a stability date of 2028 for Edition 1.1. Edition 2 has not been published, so build to Edition 1.1.
How do we decide the software safety class?
From the risk analysis. Class A: the software cannot contribute to a hazardous situation, or the risk is acceptable once risk control measures outside the software are taken into account. Class B: it can contribute to an unacceptable risk and the possible harm is non-serious injury. Class C: the possible harm is death or serious injury. Hardware, an independent software system or clinical procedures can be external risk controls, and adding them can lower the class, with the rationale kept in the risk management file.
Does IEC 62304 apply to a mobile app or other standalone software?
Yes. It applies to software that is itself a medical device as well as to software embedded in one. It does not cover validation and final release of the device, even when the device is entirely software; for standalone health software those system-level activities are in IEC 82304-1. MDA still wants the validation evidence, tested in-house and in the user’s environment for every hardware configuration in the labelling.
What is SOUP, and what do we need for open-source libraries?
SOUP is software of unknown provenance: software that is already developed and generally available but was not made for your device, or software previously developed without adequate records. Operating systems, open-source libraries, SDKs and AI frameworks usually count. Every class needs each SOUP item identified by title, manufacturer and a unique designator such as a version. For Class B and C you also specify the functional and performance requirements it must meet and the hardware and software it needs, and you consider its known anomalies in the risk analysis.
Our software is already sold elsewhere but was not developed to IEC 62304. Can we still use it?
Amendment 1 added a route for legacy software: software legally placed on the market and still marketed, without enough evidence that it was developed to the current version of the standard. You assess post-market feedback and the risks of continued use, run a gap analysis of the available records against the requirements, software architecture, system testing and software risk management clauses, close the gaps that reduce risk, and document a rationale for continued use. The minimum deliverable is software system test records.
Do we need ISO 13485 to follow IEC 62304?
IEC 62304 assumes the software is developed and maintained within a quality management system and a risk management system. For medical devices that quality system is normally ISO 13485, and the risk management system is ISO 14971, which IEC 62304 references normatively. A software company that is the legal manufacturer of a medical device will need both in any case.
Does a software update need MDA approval?
Some do. Under MDA/GD/0072, changes that affect device control, add indications, change a diagnostic or therapeutic algorithm, add or remove alarms or change the operating-system platform are significant and need approval; bug fixes and cybersecurity improvements need no notification. Until MeDC@St 3.0 launches, MDA/GD/0020 applies. IEC 62304’s maintenance, configuration management and problem resolution processes are what let you show which kind of change you made.

A software file the CAB will accept

Tell us what the software does and what documentation you have. We classify it, close the IEC 62304 gaps and write the software section of the CSDT for MDA.